GATE Instrumentation (IN): Measurements (errors and uncertainty, regression, Wheatstone, Kelvin and AC bridges, wattmeters and instrument transformers, digital voltmeters, oscilloscope probes, Lissajous figures and counters), previous-year questions solved, part 2
Part 2 of 2 of the Measurements hub: 13 more official GATE IN questions on voltage, current and power measurement, instrument transformers and digital meters, oscilloscopes, time and frequency measurement, each solved step by step and checked against the official answer key. Part 1 covers errors, uncertainty, statistics and SI units and bridges, galvanometers and resistance measurement.
Questions from official GATE IN papers (IIT Kharagpur 2022, IIT Kanpur 2023, IISc Bengaluru 2024, IIT Roorkee 2025, IIT Guwahati 2026); solutions by Skill-Lync. · Updated 19 Nov 2026 · 13 of 30 questions on this page
Balanced star of 8.2 uF capacitors, 440 V, 50 Hz; wattmeter current coil in R, pressure coil R to Y: reading?
Answer: Official key: 141.80 to 146.20
- Three equal capacitors in star on a balanced supply: the star point sits at the supply neutral. Each capacitor sees the phase voltage .
- , so , leading by .
- The current coil carries . The pressure coil is across lines R and Y, so it sees , which leads by (phase sequence RYB, assumed).
- Angle between and : .
- .
Trap: the wattmeter reads of the angle between the voltage on its pressure coil and the current in its current coil, not the load power factor. The load takes no real power at all, yet the meter reads 144.0 W. Using gives 249.4 W, and using the phase voltage gives 83.1 W. With the opposite phase sequence (RBY) the angle becomes and the meter reads , so the stated answer assumes RYB. We solved the three-phase circuit by nodal analysis with a floating star point and a near-ideal meter: 143.974 W.
Single wattmeters are often connected with the current coil in one line and the pressure coil across the other two to read reactive power in balanced three-phase systems. Getting the phase sequence and coil polarity right is part of any meter commissioning, because a reversed sequence flips the sign of such readings.
100 W wattmeter with 2:1 and 1:1 transformers, 100 ohm load, Vs = 200 V RMS: what does it read?
Answer: Official key: 0 to 0
- Trace the conducting loops. The source and the 2:1 primary form one closed loop. The 2:1 secondary, the 100 Ω resistor and the 1:1 primary form a second closed loop. The two loops are linked only magnetically.
- The current coil is joined from the lower end of the primary loop to the lower end of the secondary loop. It is the only wire between two circuits with no other conducting connection, so current has no return path. By KCL around the secondary circuit, the current-coil current is zero.
- The potential coil is across the 1:1 secondary, so it sees .
- Reading 0 W.
Trap: assuming the current coil carries the 1 A load current. That gives , exactly the full-scale value, which makes it look right. We modelled both transformers as tightly coupled inductors (50 Hz assumed) in a nodal solution: current-coil current 7.2e-16 A, potential-coil voltage 100.00 V, reading 3.1e-14 W. Moving the current coil into the 100 Ω loop gave 99.99 W.
A wattmeter reads only the current that actually flows through its current coil, so wiring checks before energising are part of every metering installation. Isolating transformers break the conducting path on purpose, and an earth link between isolated circuits carries no load current.
Multiplier with 1 ohm shunt, A1 = 1/(50 sqrt 2), +90 deg shifter; 230 V, load 230/sqrt2 (1 + j): |avg Vo|?
Answer: Official key: 2.1 to 2.5
- at . The load current is , lagging the load voltage by .
- The 1 Ω shunt carries this current: RMS.
- The attenuator gives RMS in phase with the load voltage. After the phase shifter, leads the load voltage by .
- The average of the product of two sinusoids is (RMS values), where is the angle between them. Here .
- , magnitude 2.3 V.
- The output is proportional to : the phase shifter turns this multiplier into a reactive-power meter ( var).
Trap: with a load the real and reactive powers are equal, so leaving out the phase shifter gives the same magnitude, 2.3 V, with the opposite sign. Do not take the number as proof of the method. Multiplying peak values instead of RMS values gives 4.6 V. We solved the circuit as phasors, set the source so that the load saw 230 V, generated and as time waveforms and averaged their product over one 20 ms cycle: . The sign depends on the polarity of the attenuator input, which the figure does not fix; the question asks for the magnitude.
Electronic energy meters and power analysers multiply sampled current and voltage signals; shifting the voltage by 90 degrees, digitally or with an integrator, gives reactive power in var. Utilities bill large consumers for reactive energy, so meters measure both.
Linear transducer calibrated 50 C to 150 C in and 4 mA to 20 mA out: gain in mA per degree C?
Answer: Official key: 0.15 to 0.17
- A linear transducer has a straight-line transfer: output current , with at and at .
- The gain is the slope: output span divided by input span.
- .
- The line extended back to would give , which is why the gain alone does not describe the transducer: the offset (here the 4 mA live zero at ) is the second calibration number.
Trap: dividing full-scale values, , or forgetting the 4 mA live zero, . Both fall outside the key. We sampled the straight line through the two calibration points at 11 temperatures across the span and fitted it by least squares: slope 0.1600 mA per degree C.
Most process transmitters send 4-20 mA, with 4 mA at the bottom of the calibrated range and 20 mA at the top, so a broken loop (0 mA) is easy to tell from a low reading. The PLC or DCS input card scales the current back to engineering units using the same two end points that were set when the transmitter was ranged.
Wattmeter through a PT and a CT, lagging load: alpha in terms of gamma, beta, delta and theta?
- is the phase angle between the load current and the load voltage
- is the phase angle by which pressure coil current lags the pressure coil voltage of the wattmeter
- is the phase angle between currents in the pressure coil and the current coil of the wattmeter
- is the phase angle of the voltage transformer
- is the phase angle of the current transformer
- (A)
- (B)
- (C)
- (D)
Answer: (C) Official key: C
- Take the load voltage as the reference phasor. The load current lags it by , so its angle is .
- The PT secondary voltage is displaced from by the PT phase angle , and the CT secondary current (the current-coil current) is displaced from by the CT phase angle . Whether each secondary leads or lags depends on the transformer, which is why the options carry .
- The pressure-coil current lags by (the coil inductance). With both secondaries leading: angle of , angle of .
- , which has the form : option (C).
Note: the paper does not fix a sign convention for and , so the in the options is what makes the answer convention-free. What is fixed is that and both enter with a plus sign.
Trap: subtracting (options A and D). The pressure-coil lag pulls toward the lagging load current, so the angle the wattmeter sees, , is smaller than and has to be added back. We built the phasors for 200 random cases (load angle 5 to 80 degrees, transformer phase angles up to 1.5 degrees either way, pressure-coil R and L from a range, up to 13.5 degrees), measured from the complex currents and tested each option for any fixed choice of the two signs: only (C) holds, with minus and plus when both angles are counted as the lead of the secondary over the primary.
High-voltage and high-current circuits are metered through voltage and current transformers, and metering-class instrument transformers are specified for both ratio error and phase displacement. Phase errors matter most at low power factor, where a fraction of a degree changes the measured active power noticeably, so energy meters and wattmeters on such circuits apply a phase-angle correction.
Wattmeter with pressure coil on the supply side (A) or load side (B): condition for low error in both?
- (A) for Topology-A and for Topology-B
- (B) for Topology-A and for Topology-B
- (C) for Topology-A and for Topology-B
- (D) for Topology-A and for Topology-B
Answer: (A) Official key: A
- Topology-A: the pressure coil sits on the supply side, so it sees while the current coil carries only . Reading . The extra term is the current-coil loss.
- Relative error in A . It is small when , that is a low-current load whose voltage is large compared with the coil drop.
- Topology-B: the pressure coil sits across the load, so it sees , but the current coil carries . Reading . The extra term is the pressure-coil loss.
- Relative error in B . It is small when , a heavy-current load.
- Both conditions together: (A).
Trap: swapping which loss each connection carries. Supply-side pressure coil means the current-coil drop is in the reading; load-side pressure coil means the pressure-coil current is in it. We solved both connections as circuits with the coil resistances (coils modelled as resistances, assumed) and took the reading as . With each condition set as a 1000:1 ratio, option (A) gave errors of 0.10% in both topologies, and every other option left one topology with a large error.
Power-analyser and wattmeter manuals say which way to connect the voltage terminals: on the supply side for low-current loads and on the load side for high-current loads. Compensated wattmeters add a winding that cancels the pressure-coil current in the current coil, which removes the load-side error.
Two-wattmeter method, balanced load 30 kW per phase at 0.866 pf lagging: the two wattmeter readings?
- (A)15 kW and 15 kW
- (B)22.5 kW and 7.5 kW
- (C)60 kW and 30 kW
- (D)45 kW and 45 kW
Answer: (C) Official key: C
- Total power . Power factor , so .
- The wattmeters read and , which are and .
- , so .
- Readings: 60 kW and 30 kW, option (C). Check: .
Trap: equal readings, (A) and (D), happen only at unity power factor, and (A) also treats 30 kW as the total. Taking 30 kW as the total at this power factor gives 20 kW and 10 kW, which is not offered. We built the three line voltages and currents as phasors of a balanced star load lagging by and took each reading as of its line-to-line voltage and line current: 60.00 kW and 30.00 kW, total 90.00 kW. Which meter shows 60 depends on the phase sequence.
Three-wire three-phase supplies are metered with two wattmeter elements, and many three-phase energy meters and power analysers use this two-element (Aron) connection. The ratio of the two readings also gives the power factor, and one reading turns negative once the power factor drops below 0.5.
3-1/2 digit 1.999 V DMM, 0.1% rdg + 2 digits, 100:5 class 1 CT, 0.1 ohm burden 0.5%: worst-case error in V?
of reading digits.
It is used to measure 100 A (RMS) current flowing through a line using a 100:5 ratio, Class-1 current transformer with a burden of .
The worst-case absolute error in the multimeter output is _________ V (rounded off to three decimal places).
Answer: Official key: 0.009 to 0.011
- CT secondary current: . The burden turns it into , read on the 1.999 V range.
- Class 1 CT: ratio error up to at rated current. Burden: . Worst high input: .
- On the 1.999 V range one digit is 0.001 V. The meter adds up to of reading plus .
- Worst high display: , an error of . The low side gives the same to three decimals.
- Quick form: of 0.5 V plus 2 digits 0.010 V.
Trap: combining the errors as a root sum of squares gives 0.0060 V, a probable error, not the worst case asked for. Counting only the meter specification gives 0.0025 V. We evaluated all eight corners of the CT, burden and meter tolerances: the largest deviation from 0.5 V is 0.01003 V.
Panel meters and energy meters on large currents read a CT secondary through a burden resistor or a shunt, so the system accuracy is the CT class, the burden tolerance and the meter specification together. Metering-grade CTs (class 0.2 or 0.5) are used for billing, while class 1 and below suit indication.
Switching phase-sensitive detector, V_sense 2 sin(4000 pi t), V_ref 4 sin(2000 pi t), 1 Hz PMMC: reading?
Note: Assume all components are ideal.
Answer: Official key: 0 to 0
- The comparator output is HIGH while and LOW while . So the switch passes in the first half of each reference cycle and in the second half: .
- The reference is at 1 kHz (period 1 ms); the sensed signal is at 2 kHz, so exactly one full cycle of fits in each half of the reference cycle.
- Each half-cycle of the reference therefore contains a whole sine cycle of (or of its inverse), whose average is zero.
- The PMMC movement, with a 1 Hz bandwidth, responds only to the average of : the reading is 0 V.
Trap: treating it as an ordinary rectifier and taking the full-wave mean of a 2 V sine, (reads 1). That is the reading only when the sensed signal is at the reference frequency and in phase with it; a phase-sensitive detector gives zero mean for any even harmonic of the reference, such as this 2 kHz signal (a switching detector does still respond to odd harmonics such as 3 kHz). The 4 V reference amplitude does not enter at all, it only drives the comparator. We simulated the switch and the -1 amplifier with a first-order 1 Hz meter movement for 3 s: the settled mean over the last second is below , with a ripple of 4.0 mV peak to peak.
Lock-in amplifiers and LVDT signal conditioners use this kind of synchronous switching (or multiplication) by a reference to pull a small signal at a known frequency out of noise and interference at other frequencies. The low-pass meter or filter after the switch sets how narrow the rejection band is.
3 1/2 digit DVM, accuracy +-(0.5% of reading + 1 digit), reading 10 V DC: percentage error?
Note: Accuracy of the digital voltmeter is expressed as .
- (A)
- (B)
- (C)
- (D)
Answer: (C) Official key: C
- A digit meter shows up to 1999 counts. The lowest range that holds 10 V is the 20 V range, which reads up to .
- One digit (one count) on that range is .
- Error .
- As a percentage of the reading: , option (C).
Trap: reading on the 200 V range makes one digit and the error (B). Subtracting the digit instead of adding it gives (A). We modelled a 1999-count meter that autoranges to the smallest range holding the input: it picked a 0.01 V digit and an error of 0.60% of reading.
Handheld and bench multimeter datasheets quote DC accuracy as a percentage of reading plus a number of counts for each range. The count term matters most near the bottom of a range, which is why technicians choose the lowest range that still holds the signal.
Oscilloscope with 1 MOhm input and a 10X passive probe: input resistance seen at the probe tip?
- (A)0.9
- (B)9.1
- (C)10
- (D)11
Answer: (C) Official key: C
- A 10X passive probe puts a resistor in the tip, in series with the scope input , forming a 10:1 divider.
- , so .
- Seen from the tip the two are in series: , option (C).
- A capacitor across is trimmed so that . Then the 10:1 ratio holds at all frequencies, not only at DC.
Trap: putting the 9 M tip resistor in parallel with the scope input gives 0.9 M (A), and adding a full 10 M probe to the 1 M input gives 11 (D). We found by bisection for a 10:1 ratio (9.000 M), injected a test current at the tip to get 10.000 M, and swept the compensated divider from 1 Hz to 100 MHz with an assumed 20 pF scope input: the ratio stayed at 0.1. With set 20% low the ratio drifted by 0.018 at high frequency.
Standard 10X scope probes have a 9 megohm tip resistor and an adjustable compensation capacitor, and the scope front panel has a square-wave output for trimming it. The 10 megohm input resistance loads the circuit less, but at high frequency the probe input capacitance, not the resistance, sets the loading.
Lissajous pattern with 3 horizontal and 2 vertical tangencies, X channel at 600 Hz: Y channel frequency?
- (A)300
- (B)400
- (C)600
- (D)900
Answer: (D) Official key: D
- Each time the Y signal reaches its peak the spot touches the top edge of the pattern: one horizontal tangency per Y cycle in the repeat period. Likewise each X peak touches the right edge: one vertical tangency per X cycle.
- So equals the number of horizontal tangencies divided by the number of vertical tangencies: .
- 900 Hz, option (D).
Trap: inverting the ratio gives Hz (B). We generated and (phase assumed) over one common period and counted the distinct points on the top and right edges. 900 Hz gave 3 and 2, 400 Hz gave 2 and 3, 300 Hz gave 1 and 2, 600 Hz gave 1 and 1.
Lissajous figures were the classic way to compare an unknown frequency or phase with a reference on an analogue oscilloscope. Digital scopes and frequency counters do this directly now, but the XY mode is still used to view phase relationships, such as in transformer and audio channel checks.
Counter clock 0.1 us, gated for one period of a ~100 kHz input: maximum error in measured frequency (kHz)?
Answer: () Official key: 1 to 1
- This is period measurement: the counter counts its own clock during one input period. Period , so the count is .
- The clock is not synchronised to the input, so the count is uncertain by : a relative error of in the period.
- The frequency is computed as , so it carries the same 1%: .
- Maximum error 1 kHz. A count of 99 instead of 100 reads 101.01 kHz.
Trap: stopping at the period error (0.1 s, or 1%) without converting it to a frequency error in kHz. Period mode suits low frequencies. At 100 kHz a 100-count reading is coarse, which is why counters switch to direct counting over a long gate for high frequencies. We simulated the counter: clock ticks every 0.1 s with a random phase, counted inside one input period, for inputs from 99.8 to 100.2 kHz. The largest error was 1.002 kHz, and the one-count formula gives 1.000 kHz.
Universal counters offer frequency mode (count input cycles over a gate) and period mode (count a reference clock over input cycles), and choose the better one for the signal. Reciprocal counters always measure period and invert it, averaging over many input cycles to keep resolution high at any frequency.