GATE Instrumentation (IN): Sensors and Industrial Instrumentation (strain gauges, RTDs, thermistors, thermocouples, LVDT, capacitive, piezoelectric and Hall sensors, pressure, flow and level measurement), previous-year questions solved

Every sensors and industrial instrumentation question from the last five GATE Instrumentation Engineering papers, solved: strain-gauge and RTD bridges, thermistors, thermocouples and pyrometers, capacitive, inductive, piezoelectric and Hall-effect sensors, and the pressure, flow and level transducers of a process plant. Each answer matches the official key and was recomputed by code from the sensor equation with the values printed in the paper.

Questions from official GATE IN papers (IIT Kharagpur 2022, IIT Kanpur 2023, IISc Bengaluru 2024, IIT Roorkee 2025, IIT Guwahati 2026); solutions by Skill-Lync. · Updated 19 Nov 2026 · 7 of 32 questions on this page

GATE 2022 · IN · Q.65NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges

Bridge with R1 = 1.5 k, R2 = R3 = R4 = 1 k, 10 V; 30 k switched across R1: sensitivity |dV_CD/dR1| in V/k?

In the Wheatstone bridge circuit shown, and . The switch is initially open and the voltage between the points C and D is . Upon closing the switch at , the resistance in the arm AD changes by an amount , and the voltage between C and D changes by . The sensitivity of the bridge (in volt/kiloohm), defined as , is ________ (round off to two decimal places)
Wheatstone bridge ABCD fed by a 10 V source between A (positive) and B. R1 is in arm AD, R2 in AC, R3 in DB and R4 in CB. V_CD is taken between the open terminals at C and D. A switch that closes at t = 0 in series with a 30.0 kilohm resistor is connected across R1, from A to D+−10 VSwitcht = 030.0 kΩR1 R2 R3 R4 VCD ABCD

Answer: Official key: 1.62 to 1.72

  1. A is at +10 V and B at 0 V. Node C sits on the divider , : . Node D sits on , : .
  2. Before the switch closes: .
  3. Closing the switch puts 30 kΩ in parallel with : , so .
  4. Now and , so .
  5. Sensitivity .

Trap: using the small-signal derivative at . The question defines sensitivity with the actual finite step, and 1.60 is outside the official range 1.62 to 1.72. We solved both dividers before and after the shunt and took the ratio of the two finite changes.

Where this shows up in industry

This is shunt calibration, the standard check of a strain-gauge or load-cell channel: a precision resistor switched across one bridge arm simulates a known strain, and the indicator gain is set or verified from the output step. The shunt value is chosen so that the simulated change is close to the full-scale signal.

GATE 2023 · IN · Q.51NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges

72-turn wire-wound potentiometer, 0 V to 3.6 V; wiper bridges turns 35 and 36: output voltage?

A wire-wound ‘resistive potentiometer type’ angle sensor with 72 turns is used in an application. The first turn of the potentiometer is connected to ground while its last turn is connected to 3.6 V. The width of the wiper covers two turns ensuring make-before-break. The output (wiper) voltage when the wiper is on top of both the turns 35 and 36 is _______V (rounded off to two decimal places).

Answer: (exactly with the bridged turn shorted) Official key: 1.77 to 1.78

  1. The 3.6 V is spread over 72 equal turns, so each turn drops . A wiper on turn 35 alone would read ; on turn 36 alone, .
  2. A wiper that touches turns 35 and 36 together sits between these two steps. To first order its output is the midpoint, . This is how a make-before-break wiper doubles the resolution: it adds a half step between every pair of turns.
  3. More exactly, the wiper shorts the one turn between its two contacts. That leaves 35 turns below the wiper and 36 above, 71 turns in all: .
  4. Both values give 1.77 to 1.78 V.

Trap: reading the voltage of a single turn, , which is outside the official range 1.77 to 1.78. The point of the question is that a two-turn wiper gives an output between the turn voltages. We solved the 72-resistor chain with the bridged turn shorted by nodal analysis and got 1.7746 V.

Where this shows up in industry

Wire-wound potentiometers give a staircase output, one turn per step, so their resolution is set by the number of turns; the wiper bridging two turns halves the step. Conductive-plastic tracks are used instead where a smooth, step-free position signal is needed, for example in servo feedback.

GATE 2023 · IN · Q.54NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges

Full strain-gauge bridge R = 100 ohm, dR = 1 ohm, potentiometer Rv between R_C and R_D: minimum Rv to balance?

Four strain gauges , , and , each with nominal resistance , are connected in a bridge configuration. When a force is applied, and increase by and and decrease by as shown. A potentiometer with total resistance is connected as shown. If , and , the minimum value of resistance required to balance the bridge is ______ (rounded off to two decimal places).
Strain-gauge bridge: the source V_B (positive at the top) feeds the top node; R_A = R + delta R runs from the top node to the left node and R_B = R - delta R to the right node. From the left node R_C = R - delta R goes down to one end of the potentiometer R_v; from the right node R_D = R + delta R goes down to the other end of R_v. The wiper of R_v returns to the negative side of V_B and to ground. The output V_o is taken between the right node and the left node+−VB Vo RA = R+ΔRRB = R−ΔRRC = R−ΔRRv RD = R+ΔR

Answer: Official key: 4.00 to 4.10

  1. The excitation is applied between the top node and the wiper. Let the wiper leave a part of in series with and in series with .
  2. Left half: over . Right half: over . Balance (equal output node voltages) needs .
  3. With and : , so .
  4. The wiper can only reach : , so . The smallest pot that can balance the bridge is with the wiper at the end, .
  5. .

Trap: leaving the wiper at the centre (as for an unstrained bridge) or adding to the arm. The pot must add resistance on the side of the arms that went down, , to make up for the 2 Ω imbalance on each side. We solved the balance equation for the wiper position and searched for the smallest whose wiper position stays inside .

Where this shows up in industry

Load cells and pressure sensors use four active gauges in a full bridge, two in tension and two in compression, which quadruples the output of a single gauge and cancels uniform temperature changes. The small zero offset left after bonding is trimmed with a balance resistor or potentiometer, sized like this to cover the worst expected imbalance.

GATE 2024 · IN · Q.51NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges

Full bridge of R +/- dR gauges (R = 4.5 kOhm, dR = 1.5 kOhm) read by an INA: resistance between A and B?

In the figure shown, kΩ, kΩ, and INA is assumed to be ideal. The equivalent resistance between A and B is __________ kΩ (rounded off to nearest integer).
A Wheatstone bridge of four variable resistors excited by V_in across its top and bottom nodes: R + delta R top left, R - delta R top right, R - delta R bottom left, R + delta R bottom right. The right mid-node is terminal A, wired to the plus input of an instrumentation amplifier INA; the left mid-node is terminal B, wired to its minus input+−VinR + ΔRR − ΔRR − ΔRR + ΔRAB+−INAVCC−VEEVoutINA: Instrumentation Amplifier

Answer: Official key: 4 to 4

  1. The equivalent (Thevenin) resistance between A and B is found with the ideal source replaced by a short, which joins the top and bottom nodes of the bridge. The ideal INA draws no input current, so it does not load A or B.
  2. Node A (right mid-node) now connects to the shorted top-bottom node through kΩ and kΩ in parallel: kΩ.
  3. Node B (left mid-node) does the same through and : another 2 kΩ.
  4. The two are in series between A and B: .

Trap: ignoring gives kΩ, which rounds to 5. The parallel combination of and is , smaller than . We solved the two-node network with a 1 mA test current from B to A: V.

Where this shows up in industry

The bridge output resistance sets the noise and the input bias current error at the amplifier, which is why strain-gauge full bridges are read by instrumentation amplifiers with very high input impedance and low bias current. A 350 ohm bridge, the usual load-cell value, has an output resistance of about 350 ohm.

GATE 2025 · IN · Q.53NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges

Three-wire strain gauge, GF 2, RN 125 Ohm, strain 2e-3, leads 5, 5, 4.95 Ohm: measured strain?

A metallic strain-gauge (SG) with resistance is connected as shown in the figure, where , , represent the lead wire resistances. The SG has a gauge factor of 2 and nominal resistance of 125 Ω. When the SG is subjected to a tensile strain of , the resulting change in is . The value is measured as . The and are the equivalent resistances measured between the terminals 1 and 2, and terminals 2 and 3, respectively.
If , and , the measured value of tensile strain is __________ (rounded off to two decimal places).
A strain gauge R_SG = R_N + delta R in series with a dummy resistor R_N. The top of R_N, the junction, and the bottom of R_SG each run through a lead resistance R_L1, R_L2, R_L3 (circled as the equivalent of the lead wires) to terminals 1, 2 and 3. R_EQ1 is measured between terminals 1 and 2, R_EQ2 between 2 and 3RNRSG = (RN + ΔR)1RL12RL23RL3Equivalent representationfor lead wire resistanceREQ1REQ2

Answer: Official key: 1.75 to 1.85

  1. True change: , so .
  2. Between terminals 1 and 2 (terminal 3 open): .
  3. Between terminals 2 and 3: .
  4. , so .

Trap: answering the applied strain, . The three-wire scheme cancels lead resistance only when the leads are equal; here is 0.05 Ω short of , and that 0.05 Ω error is 10% of the 0.5 Ω signal. We built the lead network and computed the resistance between each pair of terminals with the third open: and .

Where this shows up in industry

Three-wire connections are standard for quarter-bridge strain gauges and PT100 RTDs because they cancel lead resistance when the leads match. This question shows why the leads must be the same gauge and length: a small mismatch is a direct offset when the signal is only a fraction of an ohm.

GATE 2025 · IN · Q.58NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges

Thermistor reads 2.25 kOhm at 30 C and 1.17 kOhm at 60 C: material constant beta?

The resistance of a thermistor is measured to be 2.25 kΩ at 30 °C and 1.17 kΩ at 60 °C. Its material constant is __________ K (rounded off to two decimal places).

Answer: Official key: 2160 to 2210

  1. An NTC thermistor follows with in kelvin.
  2. Taking the ratio of the two readings: .
  3. K, K: , and .
  4. (with 273 instead of 273.15, 2199.4 K; both are in the key range).

Trap: putting the temperatures in degrees Celsius gives , nonsense for a material constant. We checked that returns 1.17 kΩ exactly with this .

Where this shows up in industry

The beta value is printed on every NTC thermistor datasheet, typically between about 3000 and 4500 K for common parts, and firmware uses the beta equation to convert the measured resistance to temperature in battery packs, chargers and HVAC sensors.

GATE 2026 · IN · Q.25MCQ · 1 markResistive sensors: strain gauges, RTDs, thermistors and bridges

Two active gauges in a half bridge on a cantilever, 10 V supply, 1 mV output, GF 2.5: strain in microstrain?

Active strain gauges are connected in a half-bridge configuration to measure strain in a cantilever element. When powered by a voltage source of , the voltage at the output of the bridge is . If the gauge factor is , the strain in the cantilever element is ________ microstrain.
  1. (A)40
  2. (B)80
  3. (C)20
  4. (D)160

Answer: (B) Official key: B

  1. In a half bridge two active gauges sit in adjacent arms. On a cantilever one is bonded on the top face (tension, ) and one on the bottom face (compression, ). The other half of the bridge is two fixed resistors .
  2. Output: . The two changes add, and this result is exactly linear.
  3. 80 microstrain: option (B).

Trap: the quarter-bridge formula gives 160 microstrain (D), and the full-bridge formula gives 40 (A). Count the active arms before choosing the factor. We solved the bridge exactly for a 1 mV output: two active gauges gave 80.00 microstrain, one gauge 160.0 and four gauges 40.0.

Where this shows up in industry

Bending-beam load cells and cantilever force sensors bond gauges on opposite faces so one sees tension and the other compression; in adjacent bridge arms they double the output and an equal temperature-driven resistance change in both cancels. This is why load-cell sensitivity is quoted in mV per V of excitation.

From GATE prep to a instrumentation, IoT or embedded career

Talk to a counsellor.

Talk to a career counsellor

Free 20-minute call. We'll map the right course to where you want to work.

By submitting you agree to our Terms and Privacy Policy.

Talk to a career counsellor

Talk to a career counsellor

Free 20-minute call. We'll map the right course to where you want to work.

By submitting you agree to our Terms and Privacy Policy.