Every sensors and industrial instrumentation question from the last five GATE Instrumentation Engineering papers, solved: strain-gauge and RTD bridges, thermistors, thermocouples and pyrometers, capacitive, inductive, piezoelectric and Hall-effect sensors, and the pressure, flow and level transducers of a process plant. Each answer matches the official key and was recomputed by code from the sensor equation with the values printed in the paper.
Questions from official GATE IN papers (IIT Kharagpur 2022, IIT Kanpur 2023, IISc Bengaluru 2024, IIT Roorkee 2025, IIT Guwahati 2026); solutions by Skill-Lync. · Updated 19 Nov 2026 · 7 of 32 questions on this page
GATE 2022 · IN · Q.65NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges
Bridge with R1 = 1.5 k, R2 = R3 = R4 = 1 k, 10 V; 30 k switched across R1: sensitivity |dV_CD/dR1| in V/k?
In the Wheatstone bridge circuit shown, R1=1.5kΩ and R2=R3=R4=1kΩ. The switch is initially open and the voltage between the points C and D is VCD. Upon closing the switch at t=0, the resistance in the arm AD changes by an amount δR1, and the voltage between C and D changes by δVCD. The sensitivity of the bridge (in volt/kiloohm), defined as δR1δVCD, is ________ (round off to two decimal places)
Answer: ∣δVCD/δR1∣=1.65V/kΩOfficial key: 1.62 to 1.72
A is at +10 V and B at 0 V. Node C sits on the divider R2,R4:VC=10×1+11=5V. Node D sits on R1,R3:VD=10×R1+R3R3=10×2.51=4V.
Before the switch closes: VCD=5−4=1.000V.
Closing the switch puts 30 kΩ in parallel with R1:R1′=31.51.5×30=1.4286kΩ, so δR1=−0.0714kΩ.
Now VD=10×2.42861=4.1176V and VCD=0.8824V, so δVCD=−0.1176V.
Sensitivity =−0.0714−0.1176=1.65V/kΩ.
Trap: using the small-signal derivative (R1+R3)210R3=1.60V/kΩ at R1=1.5kΩ. The question defines sensitivity with the actual finite step, and 1.60 is outside the official range 1.62 to 1.72. We solved both dividers before and after the shunt and took the ratio of the two finite changes.
Where this shows up in industry
This is shunt calibration, the standard check of a strain-gauge or load-cell channel: a precision resistor switched across one bridge arm simulates a known strain, and the indicator gain is set or verified from the output step. The shunt value is chosen so that the simulated change is close to the full-scale signal.
GATE 2023 · IN · Q.51NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges
72-turn wire-wound potentiometer, 0 V to 3.6 V; wiper bridges turns 35 and 36: output voltage?
A wire-wound ‘resistive potentiometer type’ angle sensor with 72 turns is used in an application. The first turn of the potentiometer is connected to ground while its last turn is connected to 3.6 V. The width of the wiper covers two turns ensuring make-before-break. The output (wiper) voltage when the wiper is on top of both the turns 35 and 36 is _______V (rounded off to two decimal places).
Answer: Vo≈1.775V (exactly 1.7746V with the bridged turn shorted)Official key: 1.77 to 1.78
The 3.6 V is spread over 72 equal turns, so each turn drops 3.6/72=0.05V. A wiper on turn 35 alone would read 35×0.05=1.75V; on turn 36 alone, 1.80V.
A wiper that touches turns 35 and 36 together sits between these two steps. To first order its output is the midpoint, 21.75+1.80=1.775V. This is how a make-before-break wiper doubles the resolution: it adds a half step between every pair of turns.
More exactly, the wiper shorts the one turn between its two contacts. That leaves 35 turns below the wiper and 36 above, 71 turns in all: Vo=3.6×7135=1.7746V.
Both values give 1.77 to 1.78 V.
Trap: reading the voltage of a single turn, 1.75V, which is outside the official range 1.77 to 1.78. The point of the question is that a two-turn wiper gives an output between the turn voltages. We solved the 72-resistor chain with the bridged turn shorted by nodal analysis and got 1.7746 V.
Where this shows up in industry
Wire-wound potentiometers give a staircase output, one turn per step, so their resolution is set by the number of turns; the wiper bridging two turns halves the step. Conductive-plastic tracks are used instead where a smooth, step-free position signal is needed, for example in servo feedback.
GATE 2023 · IN · Q.54NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges
Full strain-gauge bridge R = 100 ohm, dR = 1 ohm, potentiometer Rv between R_C and R_D: minimum Rv to balance?
Four strain gauges RA,RB,RC and RD, each with nominal resistance R, are connected in a bridge configuration. When a force is applied, RA and RD increase by ΔR and RB and RC decrease by ΔR as shown. A potentiometer with total resistance Rv is connected as shown. If R=100Ω, and ΔR=1Ω, the minimum value of resistance Rv required to balance the bridge is ______Ω (rounded off to two decimal places).
Answer: Rv=4.04ΩOfficial key: 4.00 to 4.10
The excitation is applied between the top node and the wiper. Let the wiper leave a part x of Rv in series with RC and Rv−x in series with RD.
Left half: RA over (RC+x). Right half: RB over (RD+Rv−x). Balance (equal output node voltages) needs RA(RD+Rv−x)=RB(RC+x).
With RA=RD=101Ω and RB=RC=99Ω:101(101+Rv−x)=99(99+x), so x=200400+101Rv.
The wiper can only reach x≤Rv:400+101Rv≤200Rv, so Rv≥99400. The smallest pot that can balance the bridge is with the wiper at the RD end, x=Rv.
Rv,min=400/99=4.04Ω.
Trap: leaving the wiper at the centre (as for an unstrained bridge) or adding Rv to the RD arm. The pot must add resistance on the side of the arms that went down, RC, to make up for the 2 Ω imbalance on each side. We solved the balance equation for the wiper position and searched for the smallest Rv whose wiper position stays inside 0≤x≤Rv.
Where this shows up in industry
Load cells and pressure sensors use four active gauges in a full bridge, two in tension and two in compression, which quadruples the output of a single gauge and cancels uniform temperature changes. The small zero offset left after bonding is trimmed with a balance resistor or potentiometer, sized like this to cover the worst expected imbalance.
GATE 2024 · IN · Q.51NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges
Full bridge of R +/- dR gauges (R = 4.5 kOhm, dR = 1.5 kOhm) read by an INA: resistance between A and B?
In the figure shown, R=4.5 kΩ, ΔR=1.5 kΩ, and INA is assumed to be ideal. The equivalent resistance between A and B is __________ kΩ (rounded off to nearest integer).
Answer: RAB=4kΩOfficial key: 4 to 4
The equivalent (Thevenin) resistance between A and B is found with the ideal source Vin replaced by a short, which joins the top and bottom nodes of the bridge. The ideal INA draws no input current, so it does not load A or B.
Node A (right mid-node) now connects to the shorted top-bottom node through R−ΔR=3 kΩ and R+ΔR=6 kΩ in parallel: 93×6=2 kΩ.
Node B (left mid-node) does the same through R+ΔR and R−ΔR: another 2 kΩ.
The two are in series between A and B: RAB=2+2=4kΩ.
Trap: ignoring ΔR gives R/2+R/2=R=4.5 kΩ, which rounds to 5. The parallel combination of R+ΔR and R−ΔR is (R2−ΔR2)/(2R), smaller than R/2. We solved the two-node network with a 1 mA test current from B to A: VA−VB=4.000 V.
Where this shows up in industry
The bridge output resistance sets the noise and the input bias current error at the amplifier, which is why strain-gauge full bridges are read by instrumentation amplifiers with very high input impedance and low bias current. A 350 ohm bridge, the usual load-cell value, has an output resistance of about 350 ohm.
GATE 2025 · IN · Q.53NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges
A metallic strain-gauge (SG) with resistance RSG is connected as shown in the figure, where RL1,RL2,RL3 represent the lead wire resistances. The SG has a gauge factor of 2 and nominal resistance RN of 125 Ω. When the SG is subjected to a tensile strain of 2×10−3, the resulting change in RSG is ΔR. The ΔR value is measured as ΔRMEAS=REQ2−REQ1. The REQ1 and REQ2 are the equivalent resistances measured between the terminals 1 and 2, and terminals 2 and 3, respectively. If RL1=RL2=5Ω, and RL3=4.95Ω, the measured value of tensile strain is __________ ×10−3 (rounded off to two decimal places).
Answer: εmeas=1.80×10−3Official key: 1.75 to 1.85
True change: ΔR=GFεRN=2×2×10−3×125=0.5Ω, so RSG=125.5Ω.
Between terminals 1 and 2 (terminal 3 open): REQ1=RL1+RN+RL2=5+125+5=135Ω.
Between terminals 2 and 3: REQ2=RL2+RSG+RL3=5+125.5+4.95=135.45Ω.
ΔRMEAS=135.45−135=0.45Ω, so εmeas=GFRNΔRMEAS=2×1250.45=1.80×10−3.
Trap: answering the applied strain, 2.00×10−3. The three-wire scheme cancels lead resistance only when the leads are equal; here RL3 is 0.05 Ω short of RL1, and that 0.05 Ω error is 10% of the 0.5 Ω signal. We built the lead network and computed the resistance between each pair of terminals with the third open: REQ1=135.00 and REQ2=135.45Ω.
Where this shows up in industry
Three-wire connections are standard for quarter-bridge strain gauges and PT100 RTDs because they cancel lead resistance when the leads match. This question shows why the leads must be the same gauge and length: a small mismatch is a direct offset when the signal is only a fraction of an ohm.
GATE 2025 · IN · Q.58NAT · 2 marksResistive sensors: strain gauges, RTDs, thermistors and bridges
Thermistor reads 2.25 kOhm at 30 C and 1.17 kOhm at 60 C: material constant beta?
The resistance of a thermistor is measured to be 2.25 kΩ at 30 °C and 1.17 kΩ at 60 °C. Its material constant β is __________ K (rounded off to two decimal places).
Answer: β=2201.43KOfficial key: 2160 to 2210
An NTC thermistor follows R=R0eβ(T1−T01) with T in kelvin.
Taking the ratio of the two readings: lnR2R1=β(T11−T21).
T1=303.15 K, T2=333.15 K: T11−T21=2.9706×10−4K−1, and ln1.172.25=0.6539.
β=2.9706×10−40.6539=2201.43K (with 273 instead of 273.15, 2199.4 K; both are in the key range).
Trap: putting the temperatures in degrees Celsius gives β=39.2, nonsense for a material constant. We checked that R1eβ(1/T2−1/T1) returns 1.17 kΩ exactly with this β.
Where this shows up in industry
The beta value is printed on every NTC thermistor datasheet, typically between about 3000 and 4500 K for common parts, and firmware uses the beta equation to convert the measured resistance to temperature in battery packs, chargers and HVAC sensors.
GATE 2026 · IN · Q.25MCQ · 1 markResistive sensors: strain gauges, RTDs, thermistors and bridges
Two active gauges in a half bridge on a cantilever, 10 V supply, 1 mV output, GF 2.5: strain in microstrain?
Active strain gauges are connected in a half-bridge configuration to measure strain in a cantilever element. When powered by a voltage source of 10V, the voltage at the output of the bridge is 1mV. If the gauge factor is 2.5, the strain in the cantilever element is ________ microstrain.
(A)40
(B)80
(C)20
(D)160
Answer: (B)Official key: B
In a half bridge two active gauges sit in adjacent arms. On a cantilever one is bonded on the top face (tension, R(1+Gε)) and one on the bottom face (compression, R(1−Gε)). The other half of the bridge is two fixed resistors R.
Output: Vo=V[2RR(1+Gε)−21]=2VGε. The two changes add, and this result is exactly linear.
Trap: the quarter-bridge formula Vo=VGε/4 gives 160 microstrain (D), and the full-bridge formula Vo=VGε gives 40 (A). Count the active arms before choosing the factor. We solved the bridge exactly for a 1 mV output: two active gauges gave 80.00 microstrain, one gauge 160.0 and four gauges 40.0.
Where this shows up in industry
Bending-beam load cells and cantilever force sensors bond gauges on opposite faces so one sees tension and the other compression; in adjacent bridge arms they double the output and an equal temperature-driven resistance change in both cancels. This is why load-cell sensitivity is quoted in mV per V of excitation.