GATE Instrumentation (IN): Sensors and Industrial Instrumentation (strain gauges, RTDs, thermistors, thermocouples, LVDT, capacitive, piezoelectric and Hall sensors, pressure, flow and level measurement), previous-year questions solved, part 3

Part 3 of 3 of the Sensors and Industrial Instrumentation hub: 12 more official GATE IN questions on thermocouples, pyrometers and temperature measurement, pressure, force, torque, displacement, velocity and acceleration and flow, level and process variables (pH, conductivity, viscosity), each solved step by step and checked against the official answer key. The other parts cover resistive sensors: strain gauges, RTDs, thermistors and bridges and capacitive, inductive, piezoelectric and Hall-effect sensors.

Questions from official GATE IN papers (IIT Kharagpur 2022, IIT Kanpur 2023, IISc Bengaluru 2024, IIT Roorkee 2025, IIT Guwahati 2026); solutions by Skill-Lync. · Updated 19 Nov 2026 · 12 of 32 questions on this page

GATE 2022 · IN · Q.19MCQ · 1 markThermocouples, pyrometers and temperature measurement

Chromel-Alumel thermocouple, identical copper wires added from T_B to T_C: how does E2 compare with E1?

The figure shows a Chromel-Alumel thermocouple, where the junction A is held at temperature , and a thermal emf is measured using an ideal voltmeter between the open ends B1 and B2, both held at temperature . Two identical copper wires are introduced between B1-C1 and B2-C2 as shown in the figure. When C1 and C2 are held at temperature , the voltmeter reads a thermal emf . Then, ________
Two circuits. Top: a Chromel-Alumel thermocouple with its junction A at temperature T_A; the Chromel and Alumel wires end at terminals B1 and B2, both at temperature T_B, and a voltmeter across B1 and B2 reads E1. Bottom: the same thermocouple with two identical copper wires added from B1 to C1 and from B2 to C2; C1 and C2 are both at temperature T_C and the voltmeter across C1 and C2 reads E2ChromelAlumelTA AB1B2TB TB Voltmeterreading E1 ChromelAlumelTA AB1B2TB TB coppercopperC1C2TC TC Voltmeterreading E2
  1. (A)
  2. (B)
  3. (C)
  4. (D)

Answer: (D) Official key: D

  1. The thermoelectric emf of a loop is the sum, over every wire, of from one end of the wire to the other, where is that wire's Seebeck coefficient. Only the end temperatures of each wire matter.
  2. Without copper: .
  3. With copper: the copper wire from C1 to B1 adds and the copper wire from B2 to C2 adds . The two wires are identical and run between the same pair of temperatures in opposite directions, so the two terms cancel.
  4. Hence : option (D). This is the law of intermediate metals: a third metal inserted into the circuit adds no emf if both its ends are at the same temperature, and here the two copper leads together behave the same way because their B ends share and their C ends share .

Trap: thinking the new junctions at change the reading. They would, if the two extension wires were different metals or if C1 and C2 were at different temperatures. We integrated temperature-dependent Seebeck coefficients around both loops for several values of and got (3.082 mV for °C, °C) every time; replacing one copper lead with constantan shifted the reading by 0.64 mV.

Where this shows up in industry

This is why a thermocouple can be wired to copper terminals and PCB traces inside a transmitter. The cold-junction compensation sensor must measure the temperature where the thermocouple metals end (T_B here, usually the terminal block), and the terminals are kept in an isothermal block so both ends sit at that temperature.

GATE 2024 · IN · Q.14MCQ · 1 markThermocouples, pyrometers and temperature measurement

Match optical pyrometer, thermocouple, strain gauge and LVDT with their principle of measurement

Match the following measuring devices with their principle of measurement.
Measuring DevicePrinciple of Measurement
(P) Optical pyrometer(I) Variation in mutual inductance
(Q) Thermocouple(II) Change in resistance
(R) Strain gauge(III) Wavelength of radiated energy
(S) Linear variable differential transformer(IV) Electromotive force generated by two dissimilar metals
  1. (A)(P) − (III), (Q) − (IV), (R) − (II), (S) − (I)
  2. (B)(P) − (IV), (Q) − (III), (R) − (II), (S) − (I)
  3. (C)(P) − (III), (Q) − (I), (R) − (IV), (S) − (II)
  4. (D)(P) − (II), (Q) − (IV), (R) − (I), (S) − (III)

Answer: (A) Official key: A

  1. An optical (disappearing-filament) pyrometer compares the brightness of a hot body at one narrow band of wavelengths, around red light, with a calibrated filament: it works on the radiated energy at a known wavelength, so P matches (III).
  2. A thermocouple produces a Seebeck EMF at the junction of two dissimilar metals whose junctions are at different temperatures: Q matches (IV).
  3. A strain gauge is a resistor whose resistance changes with strain, : R matches (II).
  4. An LVDT moves an iron core between one primary and two secondary coils, changing the mutual inductance to each secondary: S matches (I).
  5. P-III, Q-IV, R-II, S-I is option (A).

Trap: option (B) swaps the pyrometer and thermocouple. A pyrometer never touches the target and generates no junction EMF; that non-contact radiation principle is why it is used on molten metal and kilns. We checked each option against the four device-to-principle pairs; only (A) matches all four.

Where this shows up in industry

These four devices cover most of a plant's basic sensing: thermocouples for furnace and exhaust temperatures, optical and infrared pyrometers where the target is too hot or moving to touch, strain-gauge load cells in weighing systems, and LVDTs for valve stem position and gauging in quality inspection.

GATE 2024 · IN · Q.57NAT · 2 marksThermocouples, pyrometers and temperature measurement

Diode at 250 nA, Is = 1e-15 A, Id = Is exp(alpha Vd/T), gain 1 + 5R/R: change in Vout per kelvin?

In the figure shown, the diode current is given by . is the diode voltage in volts, is the absolute temperature in Kelvin, K/V, and A is the saturation current. The dc current source, opamp and the resistors are ideal, and are assumed to be temperature independent. The change in the output voltage () per Kelvin change in temperature is __________ mV (rounded off to one decimal place).
A 250 nA current source from V_DD forward-biases a diode to ground. The diode's anode drives the plus input of an op-amp. The minus input sits at the junction of R to ground and 5R to the output V_out, a non-inverting amplifierVDD250 nAR+−+VCC−VEE5RVout

Answer: Official key: 9.5 to 10.5

  1. The 250 nA source fixes the diode current, so .
  2. With treated as constant, as the question states, (at 300 K, V).
  3. The op-amp is non-inverting with gain , so .

Trap: using a gain of 5 (as if it were an inverting stage) gives mV/K, and forgetting the amplifier gives mV/K. Note that a real silicon diode at constant current has a negative coefficient, about mV/K, because itself rises steeply with temperature; this question fixes , so the slope comes out positive. We evaluated at 299.5 K and 300.5 K from the given diode law: the difference was 10.0 mV.

Where this shows up in industry

Diode and transistor junctions are the temperature sensors inside most chips and in low-cost sensor ICs. Practical designs use the difference between two base-emitter voltages at different current densities, which is proportional to absolute temperature and removes the uncertain saturation current.

GATE 2026 · IN · Q.30MSQ · 1 markThermocouples, pyrometers and temperature measurement

Thermocouple: which is correct, Seebeck effect, resistance change, linear EMF or Joule heating?

Which of the following statements is/are correct for a thermocouple?
  1. (A)It is based on the Seebeck Effect.
  2. (B)The change in resistance is measured.
  3. (C)The voltage difference between the two junctions is linearly proportional to the temperature difference between the junctions.
  4. (D)It is based on the concept of Joule Heating.

Answer: (A) Official key: A

  1. (A) is correct. Two dissimilar metals joined at two junctions at different temperatures produce an EMF, , where is the Seebeck coefficient of the pair. That is the Seebeck effect.
  2. (B) is wrong. A thermocouple is self-generating: the EMF is read with a high-impedance input and almost no current flows. Measuring a change in resistance describes an RTD or a thermistor.
  3. (C) is wrong. The Seebeck coefficient itself changes with temperature, so the EMF is only approximately linear. For type T it rises from 38.7 to 61.8 between 0 and , and the same difference gives 4.279 mV from 0 to 100, 5.010 mV from 100 to 200 and 6.010 mV from 300 to . This is why reference tables and polynomials are used.
  4. (D) is wrong. Joule heating, , is an irreversible loss that needs current. The thermocouple EMF exists with zero current. The thermoelectric effects behind it (Seebeck, with Peltier and Thomson) are reversible.
  5. Only (A) is correct.

Trap: ticking (C) because textbooks write . That is a small-range approximation, and the statement claims exact proportionality. We evaluated the NIST ITS-90 type T reference polynomial (it reproduces the published table at 100, 200, 300 and 400 ), integrated its Seebeck coefficient to confirm (A), compared equal temperature steps for (C), and modelled an open-circuit reading on a 10 M input for (B) and (D).

Where this shows up in industry

Thermocouple transmitters and PLC input cards measure millivolts at near-zero current, add the cold-junction EMF from a temperature sensor at the terminals, and linearise with the reference polynomial for the thermocouple type instead of one mV-per-degree constant. Using a single constant over a wide range is a common source of error in home-made data loggers.

GATE 2026 · IN · Q.51MCQ · 2 marksThermocouples, pyrometers and temperature measurement

Pyrometer read 1000 C assuming emissivity 0.8; true emissivity 0.7: actual temperature nearest to?

The emitted radiant energy from a piece of metal was measured using a pyrometer. The temperature was calculated to be , assuming a surface emissivity of . It was later found that the true surface emissivity was . The actual temperature of the object is nearest to ______ .
  1. (A)1043
  2. (B)958
  3. (C)1033
  4. (D)967

Answer: (A) Official key: A

  1. The pyrometer measures the total radiant energy, which by the Stefan-Boltzmann law is with in kelvin. The measured is fixed; only the emissivity used to convert it changes.
  2. with .
  3. , nearest to 1043: option (A).
  4. A lower true emissivity means the surface radiates less per unit area than assumed, so it must be hotter to give the same reading.

Trap: each distractor is one slip. Applying the ratio to degrees Celsius gives 1033.9 (C). Inverting the ratio in kelvin gives 958.2 (B), and in Celsius 967.2 (D). Radiation laws need absolute temperature. We solved numerically and evaluated all three slips.

Where this shows up in industry

Infrared thermometers and pyrometers on steel, forging and glass lines have an emissivity setting, and a wrong setting is a common error on metal surfaces, whose emissivity depends on oxidation and finish. Two-colour (ratio) pyrometers are used where emissivity is unknown or changing, because the ratio of two wavelength bands cancels it when it is the same at both.

GATE 2022 · IN · Q.12MCQ · 1 markPressure, force, torque, displacement, velocity and acceleration

Pneumatic nozzle-flapper: what small input change does it turn into a large output change?

A pneumatic nozzle-flapper system is conventionally used to convert ________
  1. (A)Small changes in flapper’s velocity to large changes in output temperature
  2. (B)Small changes in flapper’s displacement to large changes in output temperature
  3. (C)Small changes in flapper’s velocity to large changes in output pressure
  4. (D)Small changes in flapper’s displacement to large changes in output pressure

Answer: (D) Official key: D

  1. Supply air enters through a small fixed restriction (orifice) and leaves through a nozzle aimed at a flat flapper. The pressure between the two, the nozzle back pressure , is the output.
  2. The flapper gap sets the escape area of the nozzle, . The same air flow passes the orifice and the nozzle, so , which gives .
  3. depends on the gap , a displacement, not on how fast the flapper got there. As an illustration, with an assumed 0.4 mm orifice, 1.2 mm nozzle and 1.4 bar supply, moving the flapper from 20 to 60 m changes by 0.70 bar.
  4. So a few tens of micrometres of displacement give a large change in output pressure: option (D).

Trap: picking velocity. The back pressure is a static function of the gap, so the device is a displacement-to-pressure converter. Temperature plays no part. We computed for the same final gap reached slowly and quickly: the two values were identical.

Where this shows up in industry

The nozzle-flapper is the first stage of pneumatic transmitters, controllers and valve positioners. A pneumatic relay usually follows it, because the nozzle back pressure can supply only a tiny air flow, and the relay boosts it to drive the 0.2 to 1.0 bar (3 to 15 psi) signal line or a valve actuator.

GATE 2023 · IN · Q.14MCQ · 1 markPressure, force, torque, displacement, velocity and acceleration

Match LVDT, thermistor, strain gauge and diaphragm with torque, pressure, linear position and temperature

Pair the quantities (p) to (s) with the measuring devices (i) to (iv).
(i) Linear Variable Differential Transformer (LVDT)(p) Torque
(ii) Thermistor(q) Pressure
(iii) Strain gauge(r) Linear position
(iv) Diaphragm(s) Temperature
  1. (A)(i) - (r), (ii) - (s), (iii) - (q), (iv) - (p)
  2. (B)(i) - (p), (ii) - (s), (iii) - (r), (iv) - (q)
  3. (C)(i) - (r), (ii) - (s), (iii) - (p), (iv) - (q)
  4. (D)(i) - (q), (ii) - (s), (iii) - (p), (iv) - (r)

Answer: (C) Official key: C

  1. LVDT: the core moves along the axis and the differential secondary voltage is proportional to its displacement, so it measures linear position: (i) - (r).
  2. Thermistor: a semiconductor resistor whose resistance changes strongly with temperature: (ii) - (s).
  3. Diaphragm: an elastic membrane that deflects under a pressure difference. It can only be the pressure element, so (iv) - (q).
  4. That leaves torque for the strain gauge, which is correct: gauges bonded at 45° to a shaft axis measure the shear strain caused by torque: (iii) - (p).
  5. The set (r, s, p, q) is option (C).

Trap: option (A) pairs the strain gauge with pressure. Strain gauges are indeed used in pressure sensors, but only bonded to a diaphragm, and (A) then gives torque to the diaphragm, which cannot sense it. Settle the devices with only one possible measurand first. We tested each option against a table of what each device can sense and for a one-to-one pairing: only (C) passes.

Where this shows up in industry

Shaft torque is measured with four strain gauges at 45 degrees in a full bridge, read through slip rings or a telemetry module. LVDTs are used for dimensional gauging and valve-stem position feedback, and most industrial pressure transmitters use a diaphragm sensed by strain gauges or by capacitance.

GATE 2025 · IN · Q.15MCQ · 1 markPressure, force, torque, displacement, velocity and acceleration

Cantilever force transducer, S1/S2 on top and S3/S4 below: bridge arm assignment for maximum sensitivity?

In the force transducer shown in Figure (a), four identical strain gauges S1, S2, S3, and S4 are mounted on a cantilever at equal distance from its base. S1 and S2 are mounted on the top surface and S3 and S4 are mounted on the bottom surface, as shown in the Figure (a). These strain gauges are to be connected to form a Wheatstone bridge consisting of four arms A, B, C, and D, as shown in the Figure (b). From the following options, the correct order to maximize the measurement sensitivity is
Figure (a): a tapered cantilever fixed to a wall with a downward force at its free end. Strain gauges S1 and S2 sit side by side on the top surface near the base, and S3 and S4 directly below them on the bottom surface; a small section view beside it shows S1 above S3 and S2 above S4. Figure (b): a Wheatstone bridge supplied with V_s at the top and grounded at the bottom; arm A top left, B top right, D bottom left, C bottom right; output V_o is plus at the A-D junction and minus at the B-C junctionS2S1Section atthe gaugesS1S3S2S4ForceFigure (a)Vo+−VsABDCFigure (b)
  1. (A)A → S1, B → S2, C → S4, D → S3
  2. (B)A → S1, B → S4, C → S3, D → S2
  3. (C)A → S1, B → S2, C → S3, D → S4
  4. (D)A → S1, B → S4, C → S2, D → S3

Answer: (D) Official key: D

  1. A downward force at the free end bends the cantilever so the top surface is in tension and the bottom in compression. At equal distance from the base: S1 and S2 get , S3 and S4 get .
  2. Bridge output: . To make it as large as possible, arms in opposite corners (A and C, B and D) must change the same way, and adjacent arms (A and B, A and D) must change opposite ways.
  3. With S1 (tension) in arm A, arm C must also be in tension, so S2 goes in C. Arms B and D take the compression gauges: S4 in B and S3 in D. That is option (D), a full bridge giving , four times a single gauge.

Trap: every other order gives zero output. In (A) and (C) each half of the bridge (A with D, B with C) holds one tension and one compression gauge, so both halves shift by the same amount. In (B) A and D are both in tension and B and C both in compression, so neither divider ratio changes at all. Putting the tension gauges side by side in the figure does not mean adjacent arms. We computed for in every arm order: (A) 0.0000, (B) 0.0000, (C) 0.0000, (D) 0.0020.

Where this shows up in industry

Beam and S-type load cells in weighing scales and test machines are built exactly this way: two gauges in tension and two in compression in a full bridge, which also cancels temperature effects common to all four gauges.

GATE 2025 · IN · Q.22MCQ · 1 markPressure, force, torque, displacement, velocity and acceleration

Match RVDT, thermocouple, ionization gauge and strain gauge with vacuum, force, angle and temperature

Match the following sensors with their most suitable applications.
SensorApplication
PRotary Variable Differential TransformerIVacuum measurement
QThermocoupleIIForce measurement
RIonization GaugeIIIAngular displacement measurement
SStrain GaugeIVTemperature measurement
  1. (A)P – II, Q – III, R – I, S – IV
  2. (B)P – II, Q – IV, R – III, S – I
  3. (C)P – III, Q – IV, R – II, S – I
  4. (D)P – III, Q – IV, R – I, S – II

Answer: (D) Official key: D

  1. A rotary variable differential transformer is the rotary form of the LVDT: a shaped rotor changes the coupling to two secondaries with angle. P-III.
  2. A thermocouple gives a Seebeck EMF that depends on junction temperature. Q-IV.
  3. An ionization gauge ionises the few remaining gas molecules and measures the ion current, which is proportional to gas density; it works in high and ultra-high vacuum. R-I.
  4. A strain gauge bonded to an elastic element (load cell) measures force. S-II.
  5. P-III, Q-IV, R-I, S-II is option (D).

Trap: (C) gives the ionization gauge to force and the strain gauge to vacuum. Strain-gauge diaphragm sensors do measure pressure, but not high vacuum: the pressure difference is too small to strain a diaphragm, which is why ion gauges are used there. We checked each option against the four sensor-to-application pairs; only (D) matches all four.

Where this shows up in industry

Ionization gauges watch the chamber pressure on vacuum coating and semiconductor deposition systems, RVDTs report angles such as throttle and control-surface position, and strain-gauge load cells weigh everything from tanks to trucks.

GATE 2023 · IN · Q.13MCQ · 1 markFlow, level and process variables (pH, conductivity, viscosity)

Which flow meter offers the lowest resistance to flow: turbine, orifice, venturi or electromagnetic?

Which of the following flow meters offers the lowest resistance to the flow?

Note: the published paper marks option (B) with an asterisk: in the examination this option appeared as a repeated “Turbine flow meter”. The answer is the same either way.

  1. (A)Turbine flow meter
  2. (B)Orifice flow meter
  3. (C)Venturi meter
  4. (D)Electromagnetic flow meter

Answer: (D) Official key: D

  1. Resistance to flow means the permanent pressure loss the meter adds to the line.
  2. An orifice plate forces the flow through a sharp-edged hole; the jet then expands again with a large eddy loss. Most of the differential pressure it creates is never recovered.
  3. A venturi contracts the flow gradually and recovers most of the head in its diverging cone, so its permanent loss is much smaller, but not zero.
  4. A turbine meter has a rotor (and usually flow straighteners) in the stream. Spinning the rotor against bearing friction and blade drag takes energy from the fluid, so it too adds a loss.
  5. An electromagnetic flow meter is a plain lined tube of the same bore as the pipe with electrodes flush with the wall. Nothing projects into the flow, so its extra loss is essentially that of an equal length of straight pipe: option (D).

Trap: picking the venturi because it is the textbook “low-loss” differential meter. It is low loss only compared with an orifice; any meter that needs a restriction or a rotor causes some permanent loss. Our model ranks the extra loss orifice first, then venturi and turbine, and gives the electromagnetic meter none.

Where this shows up in industry

Electromagnetic flowmeters are the usual choice for slurries, pulp and wastewater because the bore is open and adds essentially no pressure drop. The liquid must be electrically conductive (a few microsiemens per centimetre is a typical minimum), so they cannot meter hydrocarbons or gases, and very pure demineralised water needs a special low-conductivity design.

GATE 2024 · IN · Q.32NAT · 1 markFlow, level and process variables (pH, conductivity, viscosity)

Float potentiometer 0-10 kOhm for 0-20 cm, Vs = 10 V with Rs = 200 Ohm, Vout = 5 V: water level?

A linear potentiometer (0 − 10 kΩ) is used to measure the water level as shown in the figure. The resistance between A and C varies linearly from 0 to 10 kΩ for a change in water level from 0 to 20 cm. The sensor is excited using a DC voltage source, V with an internal resistance, . If V, the water level is __________ cm (rounded off to one decimal place).
A float level gauge. A DC source V_S with internal resistance R_S (inside a dashed box) excites a linear potentiometer between terminal B (top) and terminal A (bottom, grounded). The wiper C is moved by a rod from a float on the water in a tank, and the output V_out is taken from the wiper C+−RSVSBACVoutWaterFloat

Answer: Official key: 10.1 to 10.3

  1. The whole 10 kΩ track sits in series with across the source, so the track current is . is read with no load on the wiper.
  2. , so .
  3. Level: .

Trap: ignoring gives exactly half the track, cm, outside the range 10.1 to 10.3. The 200 Ω drops 0.2 V, so the wiper must sit a little higher to read 5 V. We solved V by bisection on the loaded divider: cm.

Where this shows up in industry

Float-and-potentiometer level senders are still the standard fuel gauge in many vehicles and small tanks. Because the reading depends on the excitation voltage and its source resistance, such senders are usually read ratiometrically, with the ADC reference taken from the same supply.

GATE 2026 · IN · Q.26MCQ · 1 markFlow, level and process variables (pH, conductivity, viscosity)

Glass pH electrode, 60 mV per pH, 60 mV at pH 6: pH of a solution that gives -90 mV?

A glass pH electrode behaves linearly with an output change of per unit change in pH. The measurement reference is set such that a solution of pH 6 produces an output of . The pH of a solution that results in an output of is _______ .
  1. (A)3.5
  2. (B)3
  3. (C)8
  4. (D)8.5

Answer: (D) Official key: D

  1. A glass electrode follows the Nernst equation, . The slope is negative: the output voltage falls as pH rises (about per pH at ; the paper rounds it to 60 mV).
  2. Linear model through the reference point: .
  3. Set : , so 8.5: option (D).

Trap: taking the slope as per pH gives , option (A). A more negative output means a more alkaline solution. We computed the Nernst slope from , and (-59.16 mV per pH at 298.15 K), used its sign with the given 60 mV magnitude and solved for : pH 8.50.

Where this shows up in industry

Process pH analysers convert electrode millivolts with a two-point buffer calibration that sets the offset (the mV at pH 7) and the slope. The ideal slope scales with absolute temperature, so transmitters apply automatic temperature compensation, and a calibrated slope well below the ideal value is the usual sign of an ageing glass electrode.

From GATE prep to a instrumentation, IoT or embedded career

Talk to a counsellor.

Talk to a career counsellor

Free 20-minute call. We'll map the right course to where you want to work.

By submitting you agree to our Terms and Privacy Policy.

Talk to a career counsellor

Talk to a career counsellor

Free 20-minute call. We'll map the right course to where you want to work.

By submitting you agree to our Terms and Privacy Policy.