Biopotential Signal Sources, Signal Ranges and the Man-Instrument System
A free lesson from Skill-Lync's Frontiers in Biomedical Engineering course, on Grow by Skill-Lync · Last updated 16 Nov 2026 · Session: “Module 7 Session 2 - Introduction to Biomedical Instrumentation”
What will you learn in this free Frontiers in Biomedical Engineering lesson?
This part covers bioelectric, bioacoustic, biomechanical and optical signals, including pulse oximetry using red and infrared light. It gives typical amplitude and frequency ranges, such as ECG at 0.5 to 4 millivolts and EEG at 5 to 300 microvolts, and introduces the man-instrument system with stimulus, transducer and signal conditioning.
- Types of biomedical signals and where they come from
- Typical amplitude and frequency ranges of ECG, EEG and EMG
- How pulse oximetry uses red and infrared light
- Stimulus, transducer and signal conditioning stages
What comes after lesson 19?
The next free lesson is lesson 20, “Biomedical Data Acquisition, Sensors, Alarms and Instrument Characteristics” (24 min). Later in Skill-Lync's Frontiers in Biomedical Engineering course (paid): lesson 21, Module 7 Session 4 - Introduction to Biomedical Instrumentation; lesson 22, Module 8 Session 1 - Bioelectric Signals; lesson 23, Module 8 Session 2 - Bioelectric Signals. The full course has 57 lessons (20 h 27 min of video); see the course overview or its price and syllabus.
Lesson transcript
Machine transcript of the lesson audio (lightly cleaned). Technical terms may contain errors; the video is the reference.
Now, let's take up the topic on sources on biopotential signals. So, what are the sources of biopotential signals? Now, let us see a topic called as biometrics. Biometrics is a branch of science that deals with the measurement of physiological variables and parameters. Basically, biomedical signals are used primarily for extracting information on biological system under investigation. Basically, we are trying to extract the information on the biological system, which we are going to study. So, what are these bioelectric signals? These are unique to the biomedical systems. They are basically generated by nerve cells and muscle cells. Their basic source is the cell membrane potential, which is under certain conditions may be excited to generate some kind of an action potential, which we measure on the surface of the skin.
The electric field generated by the action of many of the cell constitute the total bioelectric signal. So, the most commonly measured bioelectric signals are the electrocardiogram, electromyogram, EEG, electroencephalogram, electrooculogram, and electroretinogram. Like that, we have so many biopotential signals, which are commonly measured. Other than the bioelectric signal, the next is the bioacoustic signals. The measurement of acoustic signals generated by many of the biomedical phenomena, they provide a lot of valuable information about the underlying phenomena within the human system. For examples of this are the flow of blood into the heart through the heart walls and the flow of air through the upper and the lower airways into the lungs, which will generate typical acoustic sounds.
So, if you measure these sounds, we will be in a position to monitor the conditions of the functioning of the heart or functioning of the lungs. So, the sound of the blood flow in the heart and through the heart walls generates phonocardiographic signals, which we can monitor, and we can monitor the function of blood flow into the heart and the opening and closing of the heart walls. The next type of signals is the biomechanical signals. Basically, these signals originate from some mechanical function of the biological system.
They include all types of motions and displacement signals like pressure and flow signals. For example, the movement of the chest wall in accordance with the respiratory activity is an example of this type of a signal. So, the chest wall movements creates the displacement of the chest walls, which we can measure the respiratory activity. And also, the muscle potentials which are generated due to walking or movement of any of the muscular activity will generate the electromyographic signals. These are all biomechanical signals. The next type of signals is the biochemical signals.
The signals which are obtained as a result of any chemical measurements from the living tissues or from samples analyzed in the laboratory. The examples of these are the partial pressure of carbon dioxide, partial pressure of oxygen, and concentration of various ions in the blood. The next biopotential signal is the biomagnetic signals. Basically, there are some extremely very weak magnetic fields that are being produced by various organs, such as the brain, heart, and the lungs. And the measurement of these signals provide very valuable information, which is not available with any other biosignals, such as the bioelectric signals.
For example, one typical signal is the magnetoencephalograph, which is recorded to study the functions of the brain. Other than the electroencephalogram, which is the electrical activity, the magnetoencephalograph is also being monitored, and then it provides a lot of valuable information about the physiology of the brain. The next source of biopotential signal is the bio-optic signals. Basically, these signals are generated as a result of optical functions of the biological system. They occur either naturally or induced by any other measurement process. For example, when you are trying to measure the oxygen saturation in blood, it can be estimated by measuring the transmitted or backscattered light from a tissue at different wavelengths.
So, somewhere between 680 to 900 nanometers, the red or the IR light is going to be injected, and then you can measure the volume of blood flow and the oxygen saturation in blood. Also, imaging of soft tissues using the transmitted or backscattered light in the optical window is very commonly used for imaging of the soft tissues to identify any kind of an abnormal cell growth or any kind of a tumor or a cancerous tissue which is growing up in your soft tissues. The next source of biopotential signal is the bio-impedance signal.
Basically, it gives the impedance of the tissue, which is a source of information with respect to the composition, blood distribution, and the volume. The measurement of galvanic skin response is a typical example of this kind of a signal. The galvanic skin response gives the psychological state of an individual, which can be measured by either giving a constant voltage or a current and measuring the skin impedances at some specific points to find out the psychological state of an individual. The bio-impedance signal basically is also obtained by injecting some kind of a sinusoidal current to the tissue, and we can measure the voltage drop generated across the tissue.
So, a typical example of this is the measurement of a respiration rate onto the chest electrodes of the ECG placed at the right arm and the left arm positions, and then the impedance can be measured. By means of injecting a small variation of current, and the voltage is measured, and the impedances variations will give an indication of the respiration rate of one person, which is going to be varied as the person is breathing in and breathing out, as there is going to be an impedance variation, which is going to be shown across the drop in the voltage or the current, which is going to be continuously monitored.
Now, let's see the sources of the biopotential signals, which are generated on the human body. Basically, the electroretinogram is the biopotentials of the eye. So, the source of measurement is going to be near to the eyes. The respiration parameters, it can be measured using the spirometers and other equipments, which is going to be measuring all the rate of respiration and other the lung volumes and other vital capacities and other parameters. The electrocardiogram is going to record the electrical activity of the heart. The electrodes are going to be placed near to the heart, and then the different leads are going to be monitored using the electrocardiogram.
The electromyogram basically gives the muscle potentials, and it can be measured wherever the muscle potential is of interest, and accordingly, the sensors are going to be placed near to those muscles, and then the electromyogram is going to be recorded. The pulse oximetry basically gives the oxygen saturation in blood, and also we can measure the pulse rate. Basically, to measure the pulse oximetry, there are two techniques. Either we can go with the reflection type or the transmission type. If it is going to be the transmission type, it has to be measured only at the extremities, either at the fingertips, earlobes, or it can be in the toe tips.
If it is reflection type, then any good artery has to be selected, and then the sensor and the detector has to be placed near to that artery where the blood flow is going to be there, and then we can carry out the measurement of the pulse oximetry. Basically, the electrocardiogram measures the electrical activity of the brain, so the source of measurement is going to be on the forehead, and the complete electrodes are going to be placed onto the head, and then the different channels of EEG can be recorded.
The body temperature is going to be measured either under the tongue or under the armpit. The sensor is going to be placed, and the skin temperature can be measured. The phonocardiogram records the heart sounds, so it is going to be placed onto the chest near to the heart. Onto the chest, we place the phonocardiographic sensor, and then we are going to record the various sounds generated by the heart. The blood pressure again can be invasive or non-invasive. In case it is invasive, there is going to be a pressure sensor which is going to be inserted into the artery, and the direct blood pressure can be measured.
In case of non-invasive methods, various techniques are there. It is going to be placed on the brachial artery, and then non-invasive techniques are going to be applied, and then we are going to measure the systolic and the diastolic blood pressure. Again, the blood flow, the cardiovascular system blood flow can be measured. Again, there are various techniques. There can be invasive methods and non-invasive methods. And next is the galvanic skin response. Basically, you can measure the galvanic skin response on the palm region of a person. It can be at the distal, medial points, or it can be at the thinner and the hypotenar positions.
The sensors can be placed, and then we can measure the galvanic skin response. Basically, galvanic skin response is commonly used to measure the psychological state of an individual. These are the various types of biopotential signals which are recorded on the human subject. And if you see the different characteristics of the biopotential signals as such, and we have the biopotential signal or the biopotential measurement, and it has got some range of values, and it has got some kind of frequency component in them. For example, if you take the blood flow, there is a specific range from 1 to 300 milliliters per second.
That is the flow rate. The frequency of the blood flow rate is somewhere between 0 to 20 hertz. We can use ultrasound methods or any other electromagnetic principles are also commonly used. To measure the body temperature, basically the skin temperature or the core temperature can be measured. So, the range of the temperature of the human body varies from 32 degrees to 40 degrees is the common range which we are looking at. And it is a very slow varying signals. It ranges from DC to 0.1 hertz.
And commonly used technique for measurement of the body temperature is a thermistor, or we can use a thermocouple also. To measure the blood pressure, the range considered generally is from 0 to 400 mm of Hg. The frequency range varies from DC to 50 hertz. And here again we can use the direct method or you can use the indirect method and there are alternate techniques such as the pulse transmit time based analysis where we don't use any kind of cuff based techniques especially for the mobile applications where a person is going to be mobile. The cuff based techniques will not work. So we use an alternate method to derive the blood pressure by measuring the pulse transmit time from the heart to the extremities.
Again the cardiac output the range of measurement varies from 4 to 25 liters per minute and the frequency is from DC to 20 hertz. The method of measurement is by the flick method or the dye dilution methods. Again the ECG signals the amplitude ranges from 0.5 to 4 millivolts. The frequency ranges from 0.05 to 150 hertz. The commonly used method is the skin electrodes. Again for the electroencephalogram it is a very very low voltage signal. It ranges from 5 to 300 microvolts. The frequency of interest is between 0.5 to 150 hertz. Again we use the scalp electrodes which is also a biopotential electrode similar to what is used for ECG.
Electromyographic signals are quite good in amplitude. They range from 0.1 to 5 millivolts. Even the frequency ranges from DC to 10 kilohertz. Again here we can use needle electrodes or we can use surface electrodes to measure the EMG signals. Similarly the electro retinograph has got a frequency range from DC to 50 hertz. The signal ranges from DC to 900 microvolts. It is also a very very low voltage signal and a very low frequency signal and here we again use the surface electrodes. Basically they are of contact lens type electrodes.
Similarly we can measure other parameters. Similarly we can measure other parameters like pH, PCO2 and then the respiration rate. Respiration rate again can be measured with a sensor or we can use indirect methods. The normal values ranges from 2 to 50 breaths per minute. The frequency is again a very low frequency signal. It ranges from 0.1 to 10 hertz. Again various techniques are there. We can directly measure using a sensor or we can use an impedance method to measure the respiration rate as well. Now let's take up the topic on man instrumentation system. Basically what is this man instrumentation system? What does it comprise of? Basically man instrumentation system is an overall system including the human body and the instrumentation that are required for its measurements. Basically these are set of instruments and equipment that is utilized in the measurement of multiple characteristics plus the presentation of these information are readable, interpretable to an operator or to a doctor.
And this comprises the complete instrumentation system. And this comprises the complete instrumentation system. In a man instrumentation system the human body is treated as a black box. Basically it is an unknown system. Within which there are several kinds of signals or systems that are going to be investigated. And all these systems and subsystems are interacting with each other. And they produce different kinds of signals based on the interactions. And then we are going to see this how the signals are interacting with each other. And we are going to monitor them.
And then we are going to take some kind of a decision of what these components are and what all the frequency variations, amplitude variations, what are its kind of an abnormalities which are being generated. If there are any connections in the signals which are being affected, we are going to investigate them. Basically the components of the man instrumentation systems are the subject himself and the stimulus and we have the transducer or the sensors. We have the signal conditioning. We have the display where the parameters are going to be displaced. And then we have the processing, storage and then the remote transmission and then we have the control device.
So now let's take up all these components. What is the subject? The subject is nothing but the human being on whom the measurements are going to be carried out and is referred to as the subject under study or monitoring. The next is the stimulus. Stimulus in many cases are external triggering which is required to initiate some kind of measurement onto the human subject. This stimulus can be in the form of visual, auditory, it can be tactical or it can be a direct electrical stimulation on some part of the nervous system or on the human system.
So this forms a major component of the man instrumentation system. The next is the transducer. The transducer is a device that is capable of converting the measured signal into a form or an energy that is interpretable and it can be recorded for any study and analysis. So the next part is the signal conditioning, which is the most crucial thing. Most of the signals occurred from the human body are very weak. They are very low in amplitudes and also they are embedded with a lot of noises.
And also, as I already mentioned, there are multiple subsystems interacting with each other. So you will see that the signals produced from one system will be always added with the signal from the other system also. So basically these weak signals are required to be amplified and also these signals are of very very low in frequency and they need to be filtered and then only the signals which are of interest are to be filtered and then taken and they are going to be processed and then analyzed accordingly.
So post the analysis, the interpretation has to happen. For the interpretation to happen, the data has to be displayed. Basically the data is the electrical output which is required to be converted into a form which the human can perceive them and then the human senses can interpret it. This is a very important component of the man instrumentation system. Next is the processing storage and the transmission. Basically storage is required for later and further analysis which is to be carried out on the data and also for further referring to the previous history of the data.
And transmission is always useful for any remote consultation or remote storage or any kind of an analysis which needs to be carried out at the remote location. The control device basically this forms the final component of the system. This gives the operator the flexibility of automatic control of the stimulus. So if any stimulus is to be given to the human body. So this control device takes care of it and here we are actually achieving this control device or automatic feedback of the stimulus by incorporating a feedback loop into the system.
So that is how we achieve this control of these devices. So in a measurement of physiological parameters in a man instrumentation system. Here the basic concept is to measure the physiological parameters which are of unknown quantity. But these quantities have been already been measured on different systems and they have been standardized and they are already known to us and they are well documented. Now on the human system we measure this physical quantities or physiological quantities and then we try to compare with an already known and standardized values and then we establish whether it is a normal value or an abnormal value.
The basic parameters which we are monitoring generally on a human system are body temperature, blood pressure, heart rate, pulse rate, respiration rate, blood flow and other signals like electrocardiogram, electromyogram, electroencephalogram. We interpret them based on the various frequency components, amplitude components, how the waveforms are being generated, what frequency components are there, what ranges are there. That is how the measurements are being carried out and then being interpreted on a human system. Our mark the average представля Mercury is trying to show us how the waveforms are using are