Import of Medical Devices: Licences, Forms MD-14 and MD-15, Conditions
A free lesson from Skill-Lync's Regulatory processes for medical devices course, on Grow by Skill-Lync · Last updated 16 Nov 2026 · Session: “Week 1 Session 3”
What will you learn in this free Regulatory processes for medical devices lesson?
The instructor explains chapter 5, import of medical devices. Topics include the authorised agent applying on form MD-14, the licence granted on MD-15 for five years, the five countries accepted with a free sale certificate, the 15-day notification duty, and the test licence on MD-16 and MD-17.
- Applying for an import licence on MD-14
- Free sale certificate and the five accepted countries
- Licence validity, fees and cancellation rules
- Import test licence forms MD-16 and MD-17
What comes after lesson 3?
Later in Skill-Lync's Regulatory processes for medical devices course (paid): lesson 4, Week 2 Session 1; lesson 5, Week 2 Session 2; lesson 6, Week 2 Session 3. The full course has 27 lessons (12 h 49 min of video); see the course overview or its price and syllabus.
Lesson transcript
Machine transcript of the lesson audio (lightly cleaned). Technical terms may contain errors; the video is the reference.
Welcome to the third segment of Medical Device Regulatory Processes, Chapter 5, Import of Medical Devices. As we saw in the previous segment, what are the licensing requirements for the medical devices in-house, that means inside the country. Now we are talking about import of medical devices and the licensing requirements. So it is pretty much the same, but the forms and all the processes slightly vary compared to the previous processes. So what are the rules we are going to cover in this Chapter 5? The first is the application process for the import license, then inspection of overseas manufacturing site.
This is a requirement if it is needed, the Central Licensing Authority will inspect the overseas manufacturing site also. So again, how they are going to grant the import license, what is the validity of the license, and what is the condition to be compiled by the license holder. So these are the main aspects of getting the import license. Again, what are the rules if there is a change in the constitution? So what goes for the testing license for the imported products, and how the license has been granted?
So finally, import of investigational medical device by the government hospital or statutory medical institution for the treatment of patients and import of medical device for personal use. These are the main rules which we are going to cover in the import of medical devices topic. So what is the major form to be filled in for this import license is MD-14? So the application need to be made by the authorized agent. So we saw in the definition section who is the authorized agent, and he will be representing for the foreign manufacturing site to import the product.
So again, he will be filling this MD-14 and he will be submitting along with the specified fees, which has been said in the second schedule. So all the documents specified in fourth schedule also will accompany in order to apply for the grant of import license. So this is the sample of the MD-14 form for application to issue the import license for the medical device import. So once the application has been filed in, the Central Licensing Authority can evaluate or test the incoming medical device for its compliance.
So usually the Central Licensing Authority will be testing the medical device for its compliance, and for testing those activities, the authorizing agent should pay the fees for evaluation and test. So this test will be carried out in the Central Laboratories you would like. So if at all any additional manufacturing site license need to be there, it has to be separately applied with the additional fees also. So these are again referred in the sub rule. So if an agent is requiring two or more devices from different manufacturing site abroad, so he has to apply separately.
So if needed, the Central Authority may also visit the manufacturing site for issuing the license. So again, if manufacturing site need to be visited by the Central Institute, Central Authority, so again this will be specified in the second schedule, and the documents are specified in the fourth schedule. So that the authorizing agent should bear the price for whatever testing or visit need needed by the Central Authority. Again, see again the receipt of application. So if they are doing an inspection overseas for any person or any body, so this inspection fees also shall be barred by the agent.
So that is clearly mentioned here as the applicant shall be liable to pay the fees, other the second schedule in respect to expenditure required to connection with visit of office, manufacturing side and sub rule. So the manufacturing agency may go by themselves or they can appoint any person, all the fees shall be barred by the agent. So after the inspection what happens to the Central Authority agency if it is satisfied will grant the license in the form of MD 15 or it may reject. So in case of rejection, again the applicant may appeal to the central government within a period of 45 days.
This is similar to what we have done for the inspection in the local manufacturing site also. Same can be carried out here and within 90 days of time they have to reappeal. Again, this is the import license that will be awarded in the form of MD 15. And this is another important criteria where you don't need a license to import is a free sale certificate is already issued with a medical device by the national regulatory authority of those countries. What are the countries? So these are the major countries.
Five countries are there. Then without the clinical investigation, so we can grant an import license. So if they have a free sale certificate, we can grant an import license to import those materials. Those countries which have this free sale certificate are allowed to be imported are Australia, Canada, Japan, European Union and the United States of America. So what about the other countries? So if it is not there, so it has to go through clinical investigation in India as per the under the provision of chapter 7 of these rules.
So no matter what, we have to go through this testing investigations. So once it has been like done, then a free sale certificate from that country of origin is also furnished. So where a medical device is imported from the countries other than those referred to in sub rule 3, which we showed you, which we discussed in the previous slide, a license in case of class A, class B medical device may be granted after its safety performance has been established to public safety and performance data or through clinical investigation in the country of origin.
So if the testing is done in their countries itself for class A and B, and a free sale certificate from the countries there, then we can import for class A and class B for those countries which are not mentioned in the sub rule 3. So this is a special case in the case of investigational medical devices or in vitro diagnostic devices. We need to get a permission in the form of MD 27 and MD 29 from the central licensing authority. So to import such as a medical medicine.
So we need to import medical devices with the permission letter in the form of MD 27 and 29. Without that, we cannot import an investigational medical devices or in vitro diagnostic device. So this is the form which need to go for medical devices, investigational medical devices and also in vitro medical devices. So how long it is valid? So actually this license will be valid for 5 years. Unless and until if there is a default in the fees payment or any other cancellation or surrender or any adverse effect reporting under this license, the license shall go for at least 5 years from the date of issue.
So again for retention fees after due date has a percentage of 2 so that you are fined for at least 2 percent if you default to pay their license retention fees. So this is same as what we saw for the licensing of products inside the country also. So again the same condition go for this one. If they are failing to pay the fees, the license shall be deemed to be cancelled. So what are the conditions the license has to fulfill? Again this is a very very important thing and we need to follow this whenever importing a product.
So importing a product needs a strict action in case of any adverse reaction. So that's what we are going to see here. So the license shall be produced when requested by the medical officer. Again we have to produce this license to the medical officer whenever there is a requirement. Again if there is any adverse reaction such as market withdrawal, regulatory restrictions, license, declaration of medical device not standard quality by the regulatory authority of the country of origin. On these cases within the period of 15 days we have to notify the central authority and we have to recall the product.
The license, so We need to do all these activities as a license holder. So, 15 days is the bare minimum, means two weeks. Within two weeks, if there is any adverse effect, we need to inform the central authority. So, in these cases, the agent immediately should cease to import all the materials which are found to be falling in the category 2, as mentioned below. So, after the license and the authority, we will consider this information that is given in the class 2. So, they will issue the direction what to do for that, whether we have to withdraw, we have to distribute in the limited market segment for a limited period of time, those decisions the central licensing authority will take.
So, in order to do this, we have to get a prior approval from the central authority. So, any changes we are going to make after the information, we are going to get the approval from the central authority, after which you can carry out whatever the way you want to do the import of medical devices. So, again, there is a clause here. So, whenever you want to get a prior approval for a product which is like, you have already mentioned that the serious activity has happened. So, you need to get the approval and this approval will be given within 60 days of time.
So, you need to wait until then, you have to cease every activities, until and unless the central licensing authority gives the approval within 60 days, then only you can proceed further. So, the sixth class is saying, until and unless you receive anything, you cannot take any action. So, when there is a serious incident happened during the import and the authorized agent is making some changes in order to mitigate the incident, then the authorized agent shall get a prior approval from central licensing authority. So, once he applies for this approval, he has to get back the response from the central licensing authority within 60 days.
So, the central licensing authority will intimate whether it is approved or rejected. In case, if we are not receiving anything within the stipulated time from the central licensing authority, then those changes shall be deemed to have been approved. So, you need to get the corrective action as early as possible when there is a serious incident. So, again, if your license order, there is a minor change as specified in the sixth schedule, you have to inform the central licensing authority within the period of 60 days again.
So, these changes have taken place. So, if there is any constitutional change also in the overseas, it shall also be informed by the authorized agent. So, again, any changes in terms of specification and also in terms of constitutional change, you need to inform the central licensing authority as a authorized agent. So, again, the consignment shall accompanied by invoice statement showing the name and quantity of medical device. The license shall supply the medical device for sale and offer it for sale along with package insert and user manual.
We saw what is the importance of package insert and user manual in the previous segment itself. So, same applies for import of devices also. Rule 38 condition to be complaint with the license holder. So, where the central licensing authority is satisfied that any medical device is not in conformity with the provisions of Act, rules it may issue direction that the entire batch of such medical device may not be sold or offered for sale or may recalled from the market including hospital if where it has been stocked.
So, once the central licensing authority is convinced that this is a non-confirming product, so all the stocks can be recalled and it can be stopped to be circulated in the market. That is what the complaint of the license holder should be. So, if a single batch or all the batch of the medical device is to be recalled, they shall do as per the central licensing authority's requirement. Now, if there is a change in the constitution, again, a fresh application need to be made. Again, this is also the old application still remain until the fresh license is issued.
And it's again the discretion of the central licensing authority to approve or to reject the application as per the change in the constitution. So, next is the testing for this import license. Again, it is very, very mandatory to pay the fees and do the testing for any import devices imported from the country other than India. So, the person has to undergo the test evaluation clinical investigation and he has to demonstrate the import license ability to test and evaluate demonstration. And to train this training also if at all it is necessary to the central licensing authority in the form of MD-16 accompanied by fees specified in the second schedule.
So, this is as same as how we are doing for the indigenous products medical devices. So, we need to pay the fees in order to import the medical devices as per the central licensing authority. So, once the receipt as application of the receipt is made, the investigation plan are there and the applications are submitted. The central licensing authority will determine what quantity of medical devices and how the investigation should go. Everything shall be decided by the central investigation authority. So, this is the form 16 for the license of import of medical devices for the purpose of clinical investigation.
So, again, once it is accepted, the central licensing authority will give a grant in the form of MD-17 or it may reject. So, this all the things will happen within the 30 days of submitting your application. So, this is almost same as what we do for indigenous ones. So, once the license has been given, again, this should be exclusively used for only the licensing purpose for the test and evaluation clinical investigation alone. So, again, this license is only for testing. Hence, it has to be used as per the guidance given by the central licensing authority.
So, if at all, if you are deviating from any processes, it has to be notified to the central licensing authority. So, this is the license given by the central licensing unit in the form of MD-17 for granting the application. The holder should also maintain the record of activity what they are doing. Again, whatever consignment they are taking in, they have to give complete details in the form of invoice or statement showing the name, quantity of the requirement, everything. So, this license of MD-17 issued will be valid for three years for the holder.
So, now, again, medical devices including in vitro diagnostic division, which are not used or may be permitted to be exported or destroyed under intimation of the central licensing authority. Again, so, this is a self-explanatory one. So, wherever this is not used, they will take the necessary action in the country itself. So, if there is anything contravening with the provisions of the rules against central authority, shall issue cause of notice. So, why order should not be made to cancel the license? So, you need to explain if anything contravenes with the provisions of these rules.
Again, the central licensing authority has the rights to cancel the license. Once they cancel the license, they are giving the opportunity to explain in defense for the cancellation. So, you ought to appeal to the central government within 45 days of notice of cancellation. So, this is a self-explanatory point of rule 41. So, a small quantity of investigation in medicinal device, the import of which is not allowed but approved in the country of origin may be allowed to import by the central licensing authority for the treatment of patients suffering from a life-threatening disease or disease-causing serious permanent disabilities.
This is requiring therapy for unmet medical needs. Hence, so, we have an exception here. So, central licensing authorities may allow to import in such conditions. So, again, this is done through medical officer through proper checking and everything through the form of MD 18. So, this requires also special fees separately in order to import such a medical device. So, this is the MD device. So, this is an exception where the government hospital and medical institution have for the treatment of patients. So, once this application is sent and the authority is okay with this, they will apply..
True for the treatment of patient in the form of MD-19. So again, this is also exclusively for the treatment of the patient alone. Similarly to the license which we got for testing, here it is only for the treatment. So again, we have to hold the record for all the manufactory, how much quantity is been imported, date of import, name and address of the patient, etc, etc. So again, the medical device officer has the rights to intervene anywhere by the authorization of central licensing to monitor these activities of import. Wherever needed, we have to present our records.
So again, the quantity considered necessary shall be determined by the central licensing authority. So after they are submitting the forms, they will decide what quantity should be allowed for the treatment of such conditions. So again, they can minimize also, they can increase the amount of quantity based on the medical condition. The central licensing authority has the power to do either those, whether they can reduce the amount of import or they can increase based on the scenarios. Again, we need to give the invoices and statements including all the details that comes under the import.
So here rule 43 is specifically talking about import for personal use subject to the following conditions. So this is specifically for personal use. So the medical devices shall form a part of personal baggage of a passenger and be intended for exclusive use of such passenger. Medical devices shall be declared as a personal baggage for the customer to customs authorities if they so direct. So this is for personal use actually. Special conditions where the medical device is allowed to carry with the passenger. Maybe they are in need of the device throughout their travel. So that condition such and such import license can be given.
So in those cases, the single medical use important shall not exceed the quantity specified by the registered medical practitioner. So again, this is a self-explanatory thing. So whatever prescribed by the medical devices practitioner and registered by the authorities, those only are allowed to be traveled for personal use. So again, this should have all the necessary requirement in the form of invoice statement, name inquiry and quantity of the medical devices. So again, for this we need to have a separate form of MD-20 to be applied as an application and documents confirming this bona fide personal use. So on the medical practitioners registration, all these things should be included while applying.
So this is the permission to use of medical devices for personal activities. Import of medical devices for personal use. Again, once we received by the central licensing authority, they have the authority to accept or reject. Again, accepted, then it will go MD-21. Rejected, if they are not accepted, they will be rejected within seven days of the, right? Since it's a personal use, the response is given within a week's time. Again, this is the approval form for personal use. Again, all the consignments which are being used for personal use should have all the accompanying invoices with the statement showing everything.
So what are the conditions to follow? So what conditions this need to accompany while getting the license? After getting the approval, what are the condition you need to follow? It's given in detail. So once this invoice and everything is there, then the import of medical device for personal use can be possible. So this summarizes the chapter 5 where all the licensing part has been covered. Now we will be going to chapter 6 where we talk about labeling of these medical devices. So labeling of medical devices is an important aspect of selling any medical device in the market.
So it's a mandatory requirement to ensure all the information about the medical devices has been mapped in the form of label. There are standard procedures and requirements in order to fulfill whatever has been discussed now. So let us move on to the contents which need to be covered in the labeling aspect. So labeling of medical devices. So this will be covering what are the contents which exclusively go into the labeling. And what are the things we need to do on such aspects also will be covered in rule 44.
Rule 45 will talk about the exemption where the labeling requirements of the export of medical devices need to be done. Then rule 46 talks about the unique device identification of medical device which is the new regulation that has been into practice where we use identification system which is very unique to a single part of the medical device. Even though it is from a batch, each individual medical device will have this unique device identification for maximum traceability. Then we will talk about shelf life of medical device. Then we are going to cover labeling medical devices or a new in vitro diagnostic device for the purpose of testing evaluation and clinical investigation.
So the first four lines talk about, so the following particular shall be printed in a indialable ink on a label and the shelf pack of the medical device or on a outer cover of the medical device. Under every outer covering in which the medical device is packed. So this is a requirement for all the medical device irrespective of anything, whether it is a cold storage, whether it is a normal temperature medical devices, whether it is a sterile product or it is a semi-sterile or it is under special conditions.
So we need to have this requirement on all the labeling items. So what are those? First one, name of the medical device, which goes without saying. So identifying of the medical device, which we talked about here, the unique identification system, which we will be dealing in later stage. Name of the manufacturer address of the manufacturing site, which is very, very important for the traceability. So this is an example of a medical device labeling where it consists of everything which just we spoke about. Name of the device, identification, UDI two parts and where it is being manufactured, the manufacturing address and site and the company name is given in the left hand side.
So everything has been conveyed in this label. The other information that goes into the labeling are the net quantity. So the information about the net quantity in terms of weight, measure of volumes or number of units will be correctly mentioned in the labels and also the number of device it contains in the package also expressed in the metric system of the labeling. So here we can see all the amounts. So what are the amount to be there, whether it is in a unit form or it is in a liquid form.
What is the quantity to be there? What is the lot number? All these details will go into the standard labeling of all the medical devices. So now we are going to talk about the month and year of manufacturing is very also should be included in the labels. So in case of sterile devices. So we need to mention what is the date of sterilization and how long the sterilization process and manufacturing will last. So all these things also need to go into the sterile devices. So if it is not supplied sterile, then we should also mention the date of expiry and what are the materials it is made up of.
If it is made up of any sterile materials such as stainless steel or titanium or those things also goes into the labeling system. So in this labeling we can clearly say there are indications which are mentioned in the central part in the red box where it shows about the toxicity and the dangerous caution signals which need to be incorporated in the label for any material that has hazardous substance in it or anything that is dangerous. So now what is the explanation that we can give for expiry and shelf life.
So it is very self-explanatory that the date of expiry shall be in terms of month and year and it shall mean that medical device is recommended till last day of that month and last expiry date shall proceed by the word expiry date or shelf life. So wherever required we will provide the indication when it is expired. So if you can see in the right hand side column there is a year month and a date. So another thing we need to display. Definitively mentioned is the batch number, which goes like lot number, lot or batch number or B, NO.
So, something of this sort as a standard indication should go in all the labels in order to mention the lot number. So, again, wherever it is required, we need to mention the handling conditions also. So, if it contains glass material, we have to say handle with care and other caution symbols wherever necessary. So, again, there if you are mentioning sterile material, then we have to indicate what kind of sterilization method it's been used. So, we need to give whether it is autoclave or whether it is surface sterilized by alcohol or any method which you have performed for the medical device, you need to mention the label itself.
So, again, we need to give the relevant warnings and precautions also to label so that it has been clearly indicative of the type of medical device that you are using. Again, if the device is intended for single use or multiple use is a major indication which goes into the labels of medical devices. So, if you are overprinting something like a physician sample not to be sold, it means it is a, it is a free sample. So, again, if you are not selling the product and giving us a free sample, this should be indicated in the label also.
So, now, to provide exempt from import devices, the manufacturing license number by preceding the words also need to be done. So, that the number pertains to whether it is a device that has been manufactured originally or it has been imported. So, in case of imported device where you cannot give all these details, so you need to give the license number of the importer, a name and address of the importer so that those things can be covered by name of agents. So, once it is imported, you need to give all these details in order to cover for whatever requirement that is needed in the medical devices.
This is a simple way to avoid complications on where those countries' regulation may not require those details. But in turn, we know the license number of the importer where we can take the action from there. So, again, if the next point is very self-explanatory, provided that the label may bear symbols recognized by the Bureau of Indian Standards or International Organization Stand, ISO, in lieu of the text-untext, device safety is not compromised by lack of understanding on a part of the user in case the meaning of the symbol is not obvious to the device user.
So, if the case of small-sized medical device, information cannot be printed legibly, so it can include information necessary for private identification safety such as information covered by the classes shall be included. It can be included as a separate leaflet or something like that. So, what are the exemptions of labeling requirement for export of medical devices? So, while exporting the medical devices, there are bare minimum requirements to be going into the labels which need to satisfy the regulatory aspect of India. So, what are those? So, the labels and packages or containers of device for export shall be adopted to meet a specific requirement of law of the country to which the device is to be exported.
But the following particular shall appear in an inconfigurous manner on the label of the innermost pack or self-pack of the medical device in which the device is packed and every other outer covering in which the container is packed. So, it is not a requirement for the other country which they are exporting to have all these details. Maybe the regulation of our country, we need all the requirement which we spoke in the previous slides. So, what are the bare minimum requirements are like name of the device, again the distinctive lot or batch number, date of expiry, name and address of the manufacturer, address of the actual premises where the device has been manufactured, license number.
So, again internationally recognized symbols like ISO and all these things are the bare minimum requirement for getting the exemption or export of medical devices. Even though we are going to export, all these requirements should go in the label. So, if there is a requirement for special code also as recorded by the consignee and approved by the central licensing authority, we need to incorporate those in their labels. Now, rule 46 talking about the unique device identification of medical system, which is the cutting edge, now it is a mandatory thing and it is now the technology to the core where we are tracking each individual medical devices of the same batches also.
So, for example, a barcode, a barcode, a barcode is a typical example of unique identification number where it traces back to all the raw materials and their batch numbers also. So, with effect from first day of January 2022, the medical device approved for the manufacturer, the same distribution of import shall bear unique identification, which shall contain device identifier and production identifier. So, this identifier will have all the details which is required to backtrack wherever the product is from. So, for purpose of the rule device identifier means global trade item number.
So, again this is for a general explanation, this is self-explanatory, production identifier means serial number, lot number, batch number, software, as a medical device version, manufacturing or expiry date. All these things are being traced back from the unique identification number, that is the beauty of this number, where you can track back from whichever country of origin it is from. So, this is the uniqueness of unique identified number. So, I think moving forward, all the medical device will have unique identification, device identification in order to qualify to be imported or exported.
So, this will become a universal law, already it is in place, but it will be a mandatory thing for all the medical device moving forward. Now, shelf life of the medical device, which we need to see, this is nothing but the expiry date of the medical devices. Shelf life of medical devices are very self-explanatory and what are the requirements based on the shelf life of the medical device has been explained in this rule. So, the shelf life of medical device shall be determined keeping the view that technical parameters shall ordinarily not exceed 60 months from the date of manufacture to be reckoned from month to month.
Except in these cases where satisfactory evidence is produced by the manufacturer to justify shelf life more than 60 months of a device to the satisfaction of the central licensing authority. So, anything over than 60 months, it is not considered as a shelf life. So, it falls below that. If at all you are having exceptional conditions, then you have to produce the evidence to state that it is more than 60 months of shelf life for any medical device. So, bare minimum, it cannot exceed 60 months. That is the requirement here.
Again, provided that any medical devices whose total shelf life claim is less than 90 days shall not be allowed to import by the licensing authority if it is less than 40%. So, residual shelf life on the date of import. So, if it is less than 40%, so if it is just less than 90 days, less than 90 days means, so 40% will be somewhere around 35 days. So, while it is being imported, it should not have less than 35 days of expiry. That is what they are saying here.
Similarly, so in case of shelf life device which is less than 90 days, then while we are importing, then it should not be less than 40%. That means, the date of manufacturing or the residual shelf life should not exceed 35 days. Similarly, if a product is having 90 days to 1 year as a shelf life, then the threshold is 50%. That means, it should not exceed 6 months of residual shelf life on the date of import. Similarly, for the products more than one year, it should not exceed 60%.
So, accordingly, we have to import these products. Now, labeling of medical devices or new in vitro diagnostic devices. So, any medical device or new in vitro diagnostic medical device imported or manufactured for the purpose of clinical investigation or clinical performance evaluation, test evaluation, demonstration training shall be kept in continuous bearing labels. So, this is a requirement in point of labeling for in vitro devices. So, what it contains is, it should contain name of the product. Again, the code number if it is required. Batch, yes. Serial number wherever applicable.
Date of manufacturing. Use of the date. Storage condition. Name and untrust of the manufacturer. And purpose which has been manufactured. So, for a new in vitro diagnostic device, for testing and evaluation, clinical investigation, all these things need to be satisfied. Hence, we are coming to the end of module 1 of chapter 6. So, in this module, we have covered majorly. Chapter 1 of Preliminary and Regulations of Medical Devices, where we talked about classification, grouping, essential principles, standard that has been required. And in Chapter 3, who are the licensing authorities, what are notified bodies, what are the roles and responsibilities, and also medical devices, testing centers like the accrediting body and the notified bodies.
And in Chapter 4, we are dealing in detail about the manufacturing of medical devices or sales or distribution. So, how it has been done, inspection, grant of the license, condition of the license, why it is suspended, validity, cancellation, appeal, and test license. Similarly, for import of medical devices, which are almost same to the manufacturing of medical devices, but slightly varying in terms of fees and other requirements, we saw in detail. Finally, finally, we saw the labeling requirements for a new product and also specific conditions of leading requirements in terms of shelf life.
And also, we saw what are the requirements for the new in-vitro diagnostic devices, which need to be imported.