Manufacturing Licences for Medical Devices: Forms, Audits and Timelines
A free lesson from Skill-Lync's Regulatory processes for medical devices course, on Grow by Skill-Lync · Last updated 16 Nov 2026 · Session: “Week 1 Session 2”
What will you learn in this free Regulatory processes for medical devices lesson?
The instructor covers chapter 4 of the rules: licences to manufacture for sale or distribution. It compares the application forms for Class A and B (MD3, MD4) with Class C and D (MD7, MD8), then covers audit timelines, ISO 13485, appeals and the test licence valid for three years.
- Forms MD3, MD4, MD7 and MD8 compared
- Audit before or after licence by class
- ISO 13485 quality management system checks
- Appeals, loan licences and test licences
What comes after lesson 2?
The next free lesson is lesson 3, “Import of Medical Devices: Licences, Forms MD-14 and MD-15, Conditions” (41 min). Later in Skill-Lync's Regulatory processes for medical devices course (paid): lesson 4, Week 2 Session 1; lesson 5, Week 2 Session 2; lesson 6, Week 2 Session 3. The full course has 27 lessons (12 h 49 min of video); see the course overview or its price and syllabus.
Lesson transcript
Machine transcript of the lesson audio (lightly cleaned). Technical terms may contain errors; the video is the reference.
Welcome back to the segment 2 of module 1, regulatory process for medical devices. In this section, we are going to see about chapter 4, manufacturing of medical devices for the sale or for the distribution. So, what are the contents? So, in the first tool, we are going to see about the manufacturing for sale or for the distribution of class A and B medical devices. Then, we will see the same process for class C and class D divisions. Then, the grant of manufacturing license or the loan license, which need to be follow certain requirements in order to qualify for the application.
Then, the inspection of the grant of license of class C and class D. Then, how the inspection report looks like and what is the grant of license? How the grant of license? How the grant of license takes place and the conditions for manufacturing license? So, here we are mentioning license or loan license. So, here the loan license refer to a particular manufacturer who does not have the manufacturing segment within himself. So, he will loan from the other manufacturing entity who has the facility. Hence, the term loan license.
So, what are the change in the constitution? And also, what are the types of unannounced inspection made by the state licensing authority? And whatever license they have granted, what will be the validation? What are the conditions in which the license can be suspended or canceled? Finally, we will have test license to manufacture for, of course, testing evaluation of clinical investigation. What is the conditions required for this test licensing? And in what circumstances this licenses may be canceled? So, these are the main sector of chapter 4, which we are going to see in detail in the upcoming slides.
So, first is the applications for the sale or distribution of class C and class B devices. So, preliminarily, it starts with the licensing application, which comes under form MD3 and MD4. I will be showing the sample in the upcoming slides. So, it has a fee segment installed to it and it should be filled in with the specification prescribed in the part 2 of the 4th schedule, which we will be seeing in the upcoming modules. So, one mandatory thing is they have to follow the quality management system as per Bureau of Indian Standard or as per International Standard.
This is a main requirement. What is this QMS we can see in the 5th schedule, which will be a separate module where we will be seeing in detail. So, this is the form MD3, which consists of almost all the detailing of the manufacturing location. And so, whatever details which will go into the manufacturing location and type of products that they are going to manufacture, all will be covered in this application. Again, MD4 is specifically for the loan license, where it is like in a, it's like a co-op with the facility which is already there.
So, they will be loaning those licenses from the person who is having the original manufacturing facility. So, next is once you have given the filled in forms with all the required details, the scrutiny is going to take place. Okay, so, they will be giving the results within 45 days. So, this will be the timeline for them to process your application once submitted. From the day you are submitted, all the reply from the body will be given within 45 days of acceptance or rejection. Since this is a class A and B, so, they are not going to give any audit before granting the license because the risk is a little low.
The regulation states that it is okay to move forward again. So, if they want to see also they can see but the regulation states that there is no requirement for an audit. Maybe, if they want to see, they can see but there is no requirement as such. So, once they have granted the license, then they will come for an audit within 120 days. So, this is the requirement that shall comply in order to fulfill the application process. So, once the application is granted, they are going to come for an audit within 120 days and they will approve it.
So, again, this is the MD5 form sample and 6 for your reference. So, the audit has to take place in within 120 days. So, what will happen during the audit? So, in the audit, what will be seen as priority? So, predominantly, the audit will check the conformity of the requirement of quality management system. So, whatever quality management system the manufacturer is following has to comply with the standard. So, in terms of medical devices, ISO standards are specifically there. One and only ISO standard for medical devices quality management system is ISO 13485.
So, they will definitely check for their compliance in order to qualify for the license. So, the rule 13 actually specifies the roles of the notified body. Again, we saw that in the previous segment, what are the roles that notified body has to follow and what are the requirements they are going through. So, definitely quality management system, they will be checking the standards and all the procedures and SOPs which should meet the requirement. So, next is the audit of the manufacturing for each I carried out within 35 days from the date of application.
So, in case of a manufacturer or sale or distribution of class B devices, this date may vary. So, they may take 90 days of time for completion of audit. So, once they have completed this audit, what they will do is, of course, the audit has to be transparent, they will furnish the report and send it to the state licensing authority. So, once it satisfies, then the state licensing authority will grant the license after scrutiny. So, once the notified body sends the report, the state licensing authority has to give the yes or no decision within 20 days.
So, once the state licensing authority approves it, the notified body will release the certificate. So, the notified body only recommends for certification. Once the state licensing authority approves it, then it just goes for certification. So, if it is not approved, if the license is being rejected for certain reasons, again, we can appeal it within 45 days and after the inquiry, it will be disposed within 60 days period. So, within 45 days, if the certification is rejected and we can give the application, again, again, the authorities will revert back in 60 days time.
So, once this appeal has been posted, definitely the central licensing authority or the state licensing authority will inspect the manufacturing devices for any non-conformance or any poor standard qualities according to the provision of the fifth schedule are complying or not complied. So, they will definitely check for the conformance of this regulation. So, how we are going to place the application? So, it is now an online method. Previously, it was offline. Now, it has been online for both a loan license and a normal license. So, the forms will be MD7 and MD8.
So, this is the application which is done online for class C and D devices. So, previously, we saw the whole process for class A and B. Now, this slide is specifically talking about the application of class C and D, where MD7 and MD8 are followed respectively. So, here, there are slight differences from what we saw for class B and C. So, what are the differences? So, the primary difference is once the application is presented, the central licensing authority will seek the service of an expert in the relevant field.
So, any products of class C and D is being submitted, those products experts will be used for scrutiny. He can be from an education. If you are from a company or somebody who is an expert in that product will have a go-through on the application of class C and D. This shall be done on a priority basis. So, how long it will take? So, within 45 days of application, the scrutiny need to be made. So, what they will do is immediately they will look for a person to scrutinize the documents. Once the document is scrutinized, they have to report back within 45 days. So, after that, definitely the Central Licensing Authority shall cause an inspection of the manufacturing site.
But unlike class B and A, in class C and D, there is an inspection before granting the license. So, here for inspection, definitely they will seek the help of notified body. So, representative from the notified body will go through the inspection for class C and D. So, once the notified body is through for examination, if the documents are not found to be in order, their Central Licensing Authority will themselves can cancel it electronically, saying such and such reasons. So, this is the Form 7 and 8. This will be slightly different from B and C and it has some additional details for the products which have been used.
So, what are the requirements to follow in order to qualify for this manufacturing license for all the classes? Now that we saw that class A and B, how we are going to apply, how we are going to get the license? Now, again we saw that class C and D are slightly like having a tighter scrutiny compared to class A and B. Now we will see what are the requirements which are the necessity in order to apply for any license. So, the first and foremost requirement is the Quality Management System.
The Quality Management System has to be in place during the time of audit. That is a mandatory requirement. So, this Quality Management System should be according to the Indian Bureau of Standard or International ISO Standard 13485. This Quality Management System will be the Bible for manufacturing the medical devices in any manufacturing facility. So, this process has to be according to the standard and it has to fulfill all the requirements necessary to grant a license. So, the first and foremost requirement for applying for any license or loan license is the availability of Quality Management System.
Then, who will manage this Quality Management System? To address that point, a competent technical staff. So, who is experienced enough to handle the Quality Management System will be undertaken in order to follow this process. So, basically he will be the Quality Assurance person in the manufacturing facility who will be taking care of managing the QMS. So, he needs to also have the experience, also the qualification in order to qualify as a person representing the Quality Management System. So, what are the qualification he needs? So, these are the specific qualification that it is a requirement.
So, he should have at least two years of manufacturing or testing medical devices without which he cannot qualify as a competent person for managing the QMS system. So, what are the relevant degrees? So, in branch of pharmacy, science or engineering is fair enough to be a part of a competent person with more than two years of manufacturing and testing experience. Diploma is also accepted but you need to have more than four years of experience if you are a diploma holder. Then again, so, this is more specific to the medical devices.
So, how they qualify for as a competent technical staff. So, it is more or less what we talked in the previous two slides. So, here what we discussed previously. So, before granting the license. Again, site will be inspected before period of 60 days, definitely without granting the license. Since the risk class is so high, it is a requirement that they will come for inspection. So, the inspection team must have a controlling officer. The controlling officer, we told, it is a part of an authority. He will be interested with the power to approve or reject any application of license.
He will be present as one of the members of the inspection team. Definitely, he will be there. So, the final point talks about no inspection shall be carried out without prior approval of controlling officer. So, once a controlling officer approves, the inspection will proceed. So, after the inspection, what it does? Again, it follows the same procedure as the class B and C also. The inspection report will be sent to the central licensing authority. Then, the central licensing authority will forward a copy to the same to the applicant also.
So, whatever report which has been sent to the central licensing authority, the report will also, a copy of report will also sent to the applicant regarding the inspection that has taken place before granting a license. So, again, what is the next process? Again, scrutiny. So, once it has been like granted, again, they will give the license. Again, if it is rejected within a period of 45 days from the date of inspection report, we can reapply also. So, but here, it is stating that after the application, the licensing authority should approve or reject within 45 days.
So, you will be having the results before 45, after the inspection, I think, it will be given within the time period of 45 days. So, again, it is the same thing. Once it is rejected, you can apply within 45 days of time. So, again, once you apply for the reason for rejection, the central government will reply back within 60 days. So, this is the same principle as followed for class A and B also. So, this is the MD-9 for licensing, MD-10 for loan licensing of C and D.
So, these are the application things which are in the online portal. In case of any licenses to be added after getting the license, so, we need to make a special application to grant permission to manufacture such medical devices. So, if the medical devices are in vitro diagnostic medical devices, then we need to get a prior permission from form MD-27 and or form MD-29 from Central Licensing Authority. If you are going to add any new in vitro diagnostic or investigational medical devices. So, once we got the application, if you want to add some new products, we need to follow these two points in order to add those product in your already existing management.
So, now that we obtain the license, what are the things we need to follow or complete in order to fulfill the licensing requirement. So, whenever a medical device officer asks for the license, it has to be produced now and then. So, similarly, after getting the license, they should inform the state licensing authority also of holding such a license. So, after getting the license also, we need to get the approval from the Central Licensing Authority. So, these are the intimation points which we need to give after getting the grants.
So, after that, if at all, any minor changes are there in the license that also need to be included or intimated to the state licensing or central licensing authority accordingly if you are changing anything in the license itself. So, now that you got the license, all the test batches that are released from the manufacturing facility need to be recorded and it has to be released. It means that each and every batch there should be a test procedure carried out and it should be recorded before the release of the product.
So, that whenever the inspection takes place, it should be produced at the time of audit. So, next is the license holder also need to maintain the audit or inspection book that is given in the form of MD-11. So, that notified body or medical devices officer need to record an observation or non-conformity if at all anything they come across during the inspection. So, if you are... Man... Manufacturing invasive medical devices, at least one unit of each sample from each batch need to be retained. And if you are manufacturing any in vitro diagnostic devices, so the reference purpose, it has to be maintained for 180 days after the date of expiry of such batch.
So, whenever you produce a batch, you have to retain those batchs till they are expiry and after expiry, you should hold the batch for 180 days. This is a requirement in order to go back and see whether these products are fulfilling the requirement and follow the safety and standard procedure whenever there is a serious incidence that occurs in the market. So, these, this is the auditor inspection book which needs to be mentioned and maintained in the form of MD-11. This is the sample book which, which need to be there whenever there is a inspection.
So, after requiring all these things, we need to maintain the record of all the sales that has been done and it should be transparent enough for the medical devices officers to inspect. So, whatever you have fulfilled, you need to maintain a record and it has to be shown during the time of inspection. So, this is a requirement again. So, this is an evidence that you are showing to the medical device officer that you are following the requirement as per the regulation. So, the final point is the medical device when offered for sale shall be accompanied either by its package insert or user manual wherever applicable.
So, it is the inevitable part that needs to go along with the medical devices in which all the details of the medical devices are embedded. For example, in package insert, it will have the necessary requirement like the intended purpose, intended use and procedure, materials and methods, any cautions. So, all the package insert will have the comprehensive details of the medical device, which is a requirement while selling the product. So, you need to have this user manual or package insert accordingly with the product while selling in the market.
So, all this manufacturing and testing activity shall be undertaken only under the direction and supervision of competent technical staff. So, we saw who are the competent technical staff, which were appointed previously and their qualification. So, while during the manufacturing process, there are two types of QC that has been performed by this competent technical staff. One is inline QC that is during the process of manufacturing, then final product QC. So, once the product is manufactured, then they will do some QC. So, these two things need to be done by the competitive person whenever there is a manufacturing process taking place for the medical devices.
So, that is the ultimate thing they are suggesting here. So, we need to maintain a QC check record for all these activities while performing the manufacturing. So, again, so this is for the functional type of manufacturing unit. If suppose you are stopping the manufacturing facility for 30 days, again, this has to be intimated to the central licensing authority. So, that you are notifying them again. So, this again comes to the responsibilities of a manufacturer while holding a license. So, whatever change that is happening, like now we are shutting down for 30 days, more than 30 days, we need to fortify the central licensing authority.
Change in the constitution. This is another condition where if there is any change in the constitution of the license, again, we need to intimate the central licensing authorities within 45 days. And we need to make the application again for the grant of license. So, provided that the existing license shall be deemed to be valid till such time, a fresh license is issued or application is rejected by the state and central authority as the case may be. So, this is a self-explanatory line which needs no explanation that we need to apply again as a fresh application for the license.
So, once the license is rejected, we can reappeal to the central and state government within the period of 60 days. Now, rule 28 is talking about the unannounced inspection by state licensing authority. So, the state licensing authority are deemed to take unannounced inspection also. So, they can randomly come and do the inspection. So, the second point is specifically talking about the percentage of inspection that need to be done. So, we talk about such inspection shall not be less than 2% of the total audits carried out by the notified body within the state of that class of medical device.
So, these are the two requirements when carrying out of unannounced inspection by the state licensing authority. So, it will be definitely carried out by the medical device officer. Now, coming to the rule 29, the validity of the license. So, all this validity of the license for MD-5-6, MD-9 and MD-10 will last for 5 years unless there is any cancellation or suspension by the state authority. So, if you are not having any problem so far, it will go for 5 years. So, again, if you are not paying the fees or the retention fees is delayed, so, again, they will be liable to pay a 2% of late fees.
So, again, if the fees is not properly paid, then there is all chance of the license to be cancelled. So, other conditions are if anything of the licensee is contravening with any provisions like what are the rules and the procedures that have been shared. The central licensing authorities is always liable to pass an order to suspend the license. So, it is very, very important that you have to be very careful in following the rules and regulations or whatever implied for the license. So, once it is suspended for some activities that has been recorded, we are liable to reappear within 45 days after the receipt of the copy.
So, that we can appeal for the reason for the cancellation and we can ask for an opportunity to reverse such an order. So, if the case is correct and you have all the evidences to prove that there is no error that has been committed in the manufacturing sector, the state license can revoke a suspension. So, and again, once they revoke it, they will duly publish in their website. That is, that will usually be in the CDCSO website. So, whatever we are going through shall put as transparent as possible in any forum electronically or by letter also.
So, now we will see the rule 31, test license for manufacture of, for test evaluation and clinical investigation. Now that we saw the license for the facility, now it is time to see the license required for testing purposes. So, we need to get a testing license for class A, B, C, or D that may be manufactured in the facility in the form of an application MD-12. This requires a special fees also in order to pay for the testing license. So, the application should have these documents and mainly the description of the medical device, list of the equipment's instrument and the list of the qualified person to accompany for the testing of the medical device.
So, this is the form MD-12 where the detailing of all the applicant, the competent person and the demonstration of the testing, all details will be filled in this form. So, once all the inquiry has been made and it has been satisfied, so the Central Licensing Authority will grant the testing license. So, this testing license should be maintained as a record and the quantity of the product manufactured under the test license also. So, means that whatever we are testing, we need to keep mentioning in the record wherever we are producing that much of product.
So, the testing license is valid for three years. So, this is the license for medical device manufacturing purposes. So, this will be for the clinical investigations, tester evaluation, demonstration. So, what are the conditions in which these forms are subjected to? So, this form is only for the purpose of testing and nothing else. At the specified... Place of the license. So, wherever we are getting the license for testing, the same action should be done in the same place. So, the testing shall not be done in another place where we are getting the license. So, this also gives the right for the medical device officer to intervene somewhere and investigate whatever examination we are doing as a demonstration. He has the right to come in between any testing process and ask questions.
And we need to be in a position to answer their questions whatever we claim in the test license. So, we need to also maintain all the tested material which we are stocking and keeping it for disposition. So, again, if anything is contravening on any provisions of testing, again the central licensing has already shall notice and they can cancel the testing license. Also, while canceling the central licensing authority also. While canceling the central licensing authority also gives an opportunity to appeal for defense on whatever reasons or grounds we are canceling the order. So, we need to appeal. So, this appeal need to be done within 45 days from the date they have canceled the license.
So, testing also has specific rules and regulation in order to validate its present for three years. So, again, medical devices or officers audit is available. They can intervene anytime and they can ask questions. They can cancel the license and we can also apply and prove ourselves that we are worthy of the license. So, that ends the chapter 4 and segment 2 of the module 1. We will continue with the next segment of the module 1 recording. Thanks.